| Literature DB >> 21817506 |
Masuhiro Abe1, Katsuyuki Murata, Tatsuaki Ataka, Kazuhiko Matsumoto.
Abstract
An easy calibration method based on the Langmuir adsorption theory is proposed for a carbon nanotube field-effect transistor (NTFET) biosensor. This method was applied to three NTFET biosensors that had approximately the same structure but exhibited different characteristics. After calibration, their experimentally determined characteristics exhibited a good agreement with the calibration curve. The reason why the observed characteristics of these NTFET biosensors differed among the devices was that the carbon nanotube (CNT) that formed the channel was not uniform. Although the controlled growth of a CNT is difficult, it is shown that an NTFET biosensor can be easy calibrated using the proposed calibration method, regardless of the CNT channel structures.Entities:
Year: 2008 PMID: 21817506 DOI: 10.1088/0957-4484/19/04/045505
Source DB: PubMed Journal: Nanotechnology ISSN: 0957-4484 Impact factor: 3.874