Cross-validation of input functions obtained by H₂ 15O PET imaging of rat heart and a blood flow-through detector.

Nobuyuki Kudomi1, Hannu Sipilä, Anu Autio, Vesa Oikonen, Heidi Liljenbäck, Miikka Tarkia, Jarno Laivola, Jarkko Johansson, Mika Teräs, Anne Roivainen.   

Abstract

PURPOSE: Positron emission tomography (PET) with ¹⁵O-labeled water (H₂ ¹⁵O) facilitates the visualization and quantification of blood flow in clinical investigations and also in small animals. The quantification of blood flow requires an input function, which is generally obtained by measuring radioactivity in arterial blood withdrawn during PET scanning. However, this approach is not always feasible, because abundant blood sampling may affect the physiological process being measured. The purpose of the present study was to develop and cross-validate two methods, namely, a blood- and an image-based method for obtaining the input function for blood flow studies from rat H₂ ¹⁵O PET.
METHODS: The study material consisted of two separate groups of rats. Group 1 rats were imaged twice by a high-resolution research tomograph PET camera at resting condition for a test-retest study (n = 4), and group 2 rats were imaged with and without adenosine infusion for a rest-stress study (n = 4). In group 1, radioactivity concentration in arterial blood was measured with a new flow-through detector during imaging and a blood-based input function was obtained. The image-based input function was estimated using time-activity curves from the left ventricle and myocardial regions. To validate the two input function methods, myocardial blood flow (MBF) and cerebral blood flow (CBF) were computed, and the methods were tested for reproducibility (test-retest study) and changes (rest-stress study).
RESULTS: The blood- and image-based input functions were similar, and the corresponding CBF values differed only by -6.9 ± 8.1%. In the test-retest study, both MBF and CBF showed good reproducibility, and in the rest-stress study, adenosine significantly increased both MBF (P = 0.035) and CBF (P = 0.029), compared with the resting condition.
CONCLUSION: It is possible both to measure the input function from rat arteria femoralis during H₂ ¹⁵O PET imaging and to estimate the input function from rat H₂ ¹⁵O PET images, thereby facilitating the assessment of blood flow in organs visible in PET images.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 21811861     DOI: 10.1007/s11307-011-0511-5

Source DB:  PubMed          Journal:  Mol Imaging Biol        ISSN: 1536-1632            Impact factor:   3.488


  24 in total

Review 1.  Radio-imaging in small animals.

Authors:  R Myers; S Hume; P Bloomfield; T Jones
Journal:  J Psychopharmacol       Date:  1999-12       Impact factor: 4.153

2.  Quantitative myocardial perfusion and coronary reserve in rats with 13N-ammonia and small animal PET: impact of anesthesia and pharmacologic stress agents.

Authors:  Etienne Croteau; François Bénard; M'hamed Bentourkia; Jacques Rousseau; Michel Paquette; Roger Lecomte
Journal:  J Nucl Med       Date:  2004-11       Impact factor: 10.057

3.  A femoral arteriovenous shunt facilitates arterial whole blood sampling in animals.

Authors:  Bruno Weber; Cyrill Burger; Peter Biro; Alfred Buck
Journal:  Eur J Nucl Med Mol Imaging       Date:  2002-03       Impact factor: 9.236

4.  Combinations of methods for monitoring the microcirculation of the brain.

Authors:  C Rovainen; T A Woolsey; J Dowling; D Liu; V N Semernia
Journal:  Neurosci Behav Physiol       Date:  1995 May-Jun

5.  Effects of anesthesia on functional activation of cerebral blood flow and metabolism.

Authors:  Y Nakao; Y Itoh; T Y Kuang; M Cook; J Jehle; L Sokoloff
Journal:  Proc Natl Acad Sci U S A       Date:  2001-06-05       Impact factor: 11.205

6.  In vivo quantitation of glucose metabolism in mice using small-animal PET and a microfluidic device.

Authors:  Hsiao-Ming Wu; Guodong Sui; Cheng-Chung Lee; Mayumi L Prins; Waldemar Ladno; Hong-Dun Lin; Amy S Yu; Michael E Phelps; Sung-Cheng Huang
Journal:  J Nucl Med       Date:  2007-05       Impact factor: 10.057

7.  Quantitative cerebral blood flow measurements in the rat using a beta-probe and H2 15O.

Authors:  Bruno Weber; Nicolas Späth; Matthias Wyss; Damian Wild; Cyrill Burger; Ross Stanley; Alfred Buck
Journal:  J Cereb Blood Flow Metab       Date:  2003-12       Impact factor: 6.200

8.  Modelling approach for separating blood time-activity curves in positron emission tomographic studies.

Authors:  S C Huang; J R Barrio; D C Yu; B Chen; S Grafton; W P Melega; J M Hoffman; N Satyamurthy; J C Mazziotta; M E Phelps
Journal:  Phys Med Biol       Date:  1991-06       Impact factor: 3.609

9.  Use of the left ventricular time-activity curve as a noninvasive input function in dynamic oxygen-15-water positron emission tomography.

Authors:  H Iida; C G Rhodes; R de Silva; L I Araujo; P M Bloomfield; A A Lammertsma; T Jones
Journal:  J Nucl Med       Date:  1992-09       Impact factor: 10.057

10.  Non-invasive estimation of hepatic glucose uptake from [18F]FDG PET images using tissue-derived input functions.

Authors:  N Kudomi; M J Järvisalo; J Kiss; R Borra; A Viljanen; T Viljanen; T Savunen; J Knuuti; H Iida; P Nuutila; P Iozzo
Journal:  Eur J Nucl Med Mol Imaging       Date:  2009-12       Impact factor: 9.236

View more
  1 in total

1.  Assessment of blood flow with (68)Ga-DOTA PET in experimental inflammation: a validation study using (15)O-water.

Authors:  Anu Autio; Antti Saraste; Nobuyuki Kudomi; Tiina Saanijoki; Jarkko Johansson; Heidi Liljenbäck; Miikka Tarkia; Vesa Oikonen; Hannu T Sipilä; Anne Roivainen
Journal:  Am J Nucl Med Mol Imaging       Date:  2014-09-06
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.