Accuracy and precision of hemoglobin point-of-care testing during major pediatric surgery.

N Spielmann1, J Mauch, C Madjdpour, M Schmugge, M Weiss, T Haas.   

Abstract

INTRODUCTION: The aim of the study was to compare accuracy and reproducibility of four point-of-care testing (POCT) devices (GEM® Premier 3000, ABL 800 flex, GEM® OPL™, HemoCue® B-Hemoglobin) for hemoglobin (Hb) analyzes as compared with the reference laboratory method (Sysmex XE 2100) in children undergoing major surgery.
METHODS: Pediatric patients undergoing craniofacial, spine, hip, or cancer surgery were included. Blood samples for Hb testing were taken at several intraoperative time points and generally withdrawn from the arterial catheter, if accessible.
RESULTS: A total of 256 blood samples were taken intraoperatively from 71 pediatric patients. All POCT devices showed very small bias (maximum -0.46 g/dL) to reference method as well as very good reproducibility (maximum coefficient of variation of 0.99%). However, in two cases (HemoCue), potential clinical relevant differences were observed beyond a range of 2 g/dL.
CONCLUSION: All POCT devices tested and operated by trained staff for hemoglobinometry showed reliable test results. They all allow for simple, fast, and precise bedside determination of hemoglobin concentration in the intraoperative setting.
© 2011 Blackwell Publishing Ltd.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21794097     DOI: 10.1111/j.1751-553X.2011.01363.x

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  4 in total

1.  Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

Authors:  Jan Zatloukal; Jiri Pouska; Jakub Kletecka; Richard Pradl; Jan Benes
Journal:  J Clin Monit Comput       Date:  2015-10-27       Impact factor: 2.502

Review 2.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

3.  Evaluation of a Noninvasive Hemoglobin Measurement Device to Screen for Anemia in Infancy.

Authors:  Daniel P Hsu; Alicia J French; Samuel L Madson; John M Palmer; Vinod Gidvani-Diaz
Journal:  Matern Child Health J       Date:  2016-04

4.  Simple paper-based test for measuring blood hemoglobin concentration in resource-limited settings.

Authors:  Xiaoxi Yang; Nathaniel Z Piety; Seth M Vignes; Melody S Benton; Julie Kanter; Sergey S Shevkoplyas
Journal:  Clin Chem       Date:  2013-06-20       Impact factor: 8.327

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.