Positioning accuracy of a new image-guided radiotherapy system.

Yuki Miyabe1, Akira Sawada, Kenji Takayama, Shuji Kaneko, Takashi Mizowaki, Masaki Kokubo, Masahiro Hiraoka.   

Abstract

PURPOSE: To evaluate the accuracy of the patient-positioning function of a newly developed image-guided radiotherapy system, the MHI-TM2000 (Mitsubishi Heavy Industries, Ltd., Japan).
METHODS: The isocenter positions prescribed by the lasers, MV treatment beam, and image guidance systems (kV X-ray image and kV-CBCT) were calculated using a cube phantom with a 10-mm-diameter steel ball fixed to the center of the phantom. Then, their location discrepancies were estimated. In addition, to verify the scale and orientation of the coordinate axes of the kV X-ray imaging system, positional measurements were repeated with the phantom placed at 50 mm off-isocenter along the vertical, longitudinal, and lateral directions, respectively. Further, image fusions of an anthropomorphic phantom image and the corresponding image translated by a pre-determined amount were performed.
RESULTS: The isocenter alignment among the coordinate systems was coincident within 0.5 mm in translation for the vertical, longitudinal, and lateral axes, respectively. The geometrical errors at 50 mm off-isocenter for kV X-ray images and CBCT were within 0.2 mm and 1.0 mm, respectively. The image fusion errors were within 1.0 mm in translation and 1.0 degrees in rotation, respectively. No significant difference in the image fusion accuracy was observed between the chest and pelvis phantoms.
CONCLUSIONS: The isocenter alignment among the coordinate systems was performed with high accuracy. Furthermore, the automatic image fusion function achieved sufficient patient positioning accuracy and precision for image-guided radiotherapy.

Entities:  

Mesh:

Year:  2011        PMID: 21776788     DOI: 10.1118/1.3578607

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  11 in total

1.  Effect of tumor amplitude and frequency on 4D modeling of Vero4DRT system.

Authors:  Hideharu Miura; Shuichi Ozawa; Masahiro Hayata; Shintaro Tsuda; Kiyoshi Yamada; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2017-05-05

2.  Evaluation of cone-beam computed tomography image quality assurance for Vero4DRT system.

Authors:  Hideharu Miura; Shuichi Ozawa; Masahiro Hayata; Shintarou Tsuda; Tsubasa Enosaki; Kiyoshi Yamada; Yasushi Nagata
Journal:  Rep Pract Oncol Radiother       Date:  2017-04-25

3.  Analysis of the Setup Uncertainty and Margin of the Daily ExacTrac 6D Image Guide System for Patients with Brain Tumors.

Authors:  Se An Oh; Ji Woon Yea; Min Kyu Kang; Jae Won Park; Sung Kyu Kim
Journal:  PLoS One       Date:  2016-03-28       Impact factor: 3.240

4.  On the selection of gantry and collimator angles for isocenter localization using Winston-Lutz tests.

Authors:  Weiliang Du; Jennifer L Johnson; Wei Jiang; Rajat J Kudchadker
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

5.  Dosimetric advantages of O-ring design radiotherapy system for skull-base tumors.

Authors:  Kengo Ogura; Takashi Mizowaki; Yuichi Ishida; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

6.  Commissioning and quality assurance of Dynamic WaveArc irradiation.

Authors:  Sayaka Sato; Yuki Miyabe; Kunio Takahashi; Masahiro Yamada; Mitsuhiro Nakamura; Yoshitomo Ishihara; Kenji Yokota; Shuji Kaneko; Takashi Mizowaki; Hajime Monzen; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

7.  Long-term stability assessment of a 4D tumor tracking system integrated into a gimbaled linear accelerator.

Authors:  Mami Akimoto; Mitsuhiro Nakamura; Yuki Miyabe; Nobutaka Mukumoto; Kenji Yokota; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

8.  Accuracy of image guidance using free-breathing cone-beam computed tomography for stereotactic lung radiotherapy.

Authors:  Takeshi Kamomae; Hajime Monzen; Shinichi Nakayama; Rika Mizote; Yuuichi Oonishi; Soichiro Kaneshige; Takashi Sakamoto
Journal:  PLoS One       Date:  2015-05-08       Impact factor: 3.240

9.  Effect of audio instruction on tracking errors using a four-dimensional image-guided radiotherapy system.

Authors:  Mitsuhiro Nakamura; Akira Sawada; Nobutaka Mukumoto; Kunio Takahashi; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2013-09-06       Impact factor: 2.102

10.  Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation.

Authors:  Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Nobutaka Mukumoto; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2018-04-06       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.