Initial application of a geometric QA tool for integrated MV and kV imaging systems on three image guided radiotherapy systems.

Weihua Mao1, Michael Speiser, Paul Medin, Lech Papiez, Timothy Solberg, Lei Xing.   

Abstract

PURPOSE: Several linacs with integrated kilovoltage (kV) imaging have been developed for delivery of image guided radiation therapy (IGRT). High geometric accuracy and coincidence of kV imaging systems and megavoltage (MV) beam delivery are essential for successful image guidance. A geometric QA tool has been adapted for routine QA for evaluating and characterizing the geometric accuracy of kV and MV cone-beam imaging systems. The purpose of this work is to demonstrate the application of methodology to routine QA across three IGRT-dedicated linac platforms.
METHODS: It has been applied to a Varian Trilogy (Varian Medical Systems, Palo Alto, CA), an Elekta SynergyS (Elekta, Stockholm, Sweden), and a Brainlab Vero (Brainlab AG, Feldkirchen, Germany). Both the Trilogy and SynergyS linacs are equipped with a retractable kV x-ray tube and a flat panel detector. The Vero utilizes a rotating, rigid ring structure integrating a MV x-ray head mounted on orthogonal gimbals, an electronic portal imaging device (EPID), two kV x-ray tubes, and two fixed flat panel detectors. This dual kV imaging system provides orthogonal radiographs, CBCT images, and real-time fluoroscopic monitoring. Two QA phantoms were built to suit different field sizes. Projection images of a QA phantom were acquired using MV and kV imaging systems at a series of gantry angles. Software developed for this study was used to analyze the projection images and calculate nine geometric parameters for each projection. The Trilogy was characterized five times over one year, while the SynergyS was characterized four times and the Vero once. Over 6500 individual projections were acquired and analyzed. Quantitative geometric parameters of both MV and kV imaging systems, as well as the isocenter consistency of the imaging systems, were successfully evaluated.
RESULTS: A geometric tool has been successfully implemented for calibration and QA of integrated kV and MV across a variety of radiotherapy platforms. X-ray source angle deviations up to 0.8 degrees, and detector center offsets up to 3 mm, were observed for three linacs, with the exception of the Vero, for which a significant center offset of one kV detector (prior to machine commissioning) was observed. In contrast, the gimbal-based MV source positioning of the Vero demonstrated differences between observed and expected source positions of less than 0.2 mm, both with and without gimbal rotation.
CONCLUSIONS: This initial application of this geometric QA tool shows promise as a universal, independent tool for quantitative evaluation of geometric accuracies of both MV and integrated kV imaging systems across a range of platforms. It provides nine geometric parameters of any imaging system at every gantry angle as well as the isocenter coincidence of the MV and kV image systems.

Entities:  

Mesh:

Year:  2011        PMID: 21776767     DOI: 10.1118/1.3570768

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Implementation and evaluation of an end-to-end IGRT test.

Authors:  Stephen F Kry; Jimmy Jones; Nathan L Childress
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

2.  Enhancement of megavoltage electronic portal images for markerless tumor tracking.

Authors:  Kwang-Ho Cheong; Jai-Woong Yoon; Soah Park; Taejin Hwang; Sei-Kwon Kang; Taeryool Koo; Tae Jin Han; Haeyoung Kim; Me Yeon Lee; Kyoung Ju Kim; Hoonsik Bae
Journal:  J Appl Clin Med Phys       Date:  2018-07-09       Impact factor: 2.102

3.  Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence.

Authors:  Hyejoo Kang; Rakesh Patel; John C Roeske
Journal:  J Appl Clin Med Phys       Date:  2019-09-19       Impact factor: 2.102

4.  A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery*.

Authors:  Ivan A Brezovich; Richard A Popple; Jun Duan; Sui Shen; Xingen Wu; Sidi Benhabib; Mi Huang; Rex A Cardan
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

5.  Long-term evaluation and cross-checking of two geometric calibrations of kV and MV imaging systems for Linacs.

Authors:  Tsuicheng D Chiu; Yulong Yan; Ryan Foster; Weihua Mao
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.