Literature DB >> 21776579

Mid-term results of total hip arthroplasty with the CementLess Spotorno (CLS) system.

P B de Witte1, R Brand, H G W Vermeer, H J L van der Heide, A F W Barnaart.   

Abstract

BACKGROUND: Except for those reported by the designers, there are no published mid-term results of the use of the CementLess Spotorno (CLS) Total Hip Arthroplasty system. We present the results of (1) a ten to seventeen-year follow-up prospective cohort study of this system, and (2) retrospective analyses of factors influencing clinical and radiographic outcomes.
METHODS: We studied a series of 102 consecutive CLS arthroplasties with a minimal duration of follow-up of ten years. Indications for the procedures were osteoarthritis (n = 90), rheumatoid arthritis (n = 8), and femoral head osteonecrosis (n = 4). The Merle d'Aubigné-Postel score, polyethylene wear, and radiographic status were recorded at regular intervals. Survival analyses, repeated-measures analysis of variance, and a nested case-control study (with the cases having early revision due to aseptic cup loosening within ten years after the index procedure and the controls having no early cup revision) were used for evaluation.
RESULTS: There were fourteen revisions, including nine due to aseptic cup loosening. The ten-year Kaplan-Meier survival rate was 92.2% (95% confidence interval [CI] = 86.9 to 97.5) with revision for any reason as the end point. The fifteen-year survival rate was 78.4% (95% CI = 63.9 to 92.9) with revision for any reason as the end point, 81.6% (95% CI = 66.7 to 96.5) with revision due to aseptic cup loosening as the end point, and 99.0% (95% CI = 97.0 to 100.0) with revision due to aseptic stem loosening as the end point. The average amount of polyethylene wear at the time of final follow-up was 1.92 mm (range, 0.6 to 4.3 mm). The wear rate in the cases was significantly higher than that in the controls (0.31 vs. 0.16 mm/yr, p < 0.001). Factors with a significant effect on polyethylene wear were age at surgery (a 0.3-mm increase per every ten years younger, p = 0.001) and a larger head component (an effect of 0.53 mm for the 32 vs. the 28-mm component; p < 0.0001). Male sex had an effect of -0.66 point (p = 0.07) on the final Merle d'Aubigné-Postel score.
CONCLUSIONS: The results of this CLS system, particularly with regard to the femoral stem, are comparable with those with other reliable cementless systems. Nevertheless, the prevalence of aseptic acetabular cup loosening in the second decade after the operation demonstrates a potentially substantial problem with regard to long-term survival. A high polyethylene wear rate, male sex, a younger age at the time of surgery, and a 32-mm head component size are related to inferior clinical outcomes and a higher risk of implant revision.

Entities:  

Mesh:

Year:  2011        PMID: 21776579     DOI: 10.2106/JBJS.I.01792

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  9 in total

1.  High incidence of migration with poor initial fixation of the Accolade stem.

Authors:  Craig A White; Sasha Carsen; Kevin Rasuli; Robert J Feibel; Paul R Kim; Paul E Beaulé
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  Is the bone-bonding ability of a cementless total hip prosthesis enhanced by alkaline and heat treatments?

Authors:  Kazutaka So; Ayumi Kaneuji; Tadami Matsumoto; Shuichi Matsuda; Haruhiko Akiyama
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

3.  Long-term (20- to 25-year) results of an uncemented tapered titanium femoral component and factors affecting survivorship.

Authors:  Marcus R Streit; Moritz M Innmann; Christian Merle; Thomas Bruckner; Peter R Aldinger; Tobias Gotterbarm
Journal:  Clin Orthop Relat Res       Date:  2013-05-14       Impact factor: 4.176

4.  Mid-term results of an uncemented tapered femoral stem and various factors affecting survivorship.

Authors:  Mayank Vijayvargiya; Vivek Shetty; Kiran Makwana; Harpreet Singh Suri
Journal:  J Clin Orthop Trauma       Date:  2018-02-05

5.  Outcome of the Cementless Zweymüller BICON-PLUS Cup and SL-PLUS Stem in the Very Elderly Individuals.

Authors:  Egbert J D Veen; Joost C M Schrier; Esther Van't Riet; Mark J Breslau; Alexander F W Barnaart
Journal:  Geriatr Orthop Surg Rehabil       Date:  2016-03-24

6.  Mid-term outcomes of uncemented or cemented arthroplasty revision following metal-on-metal total hip arthroplasty failure: a retrospective observational study.

Authors:  Zhao Chen; Wenli Chen; Weiguang Yu; Mingdong Zhao; Jinluan Lin; Chaoming Zhou; Hui Chen; Junxing Ye; Xianshang Zeng; Jintao Zhuang
Journal:  J Int Med Res       Date:  2020-06       Impact factor: 1.671

7.  Dimensional analysis of total hip arthroplasty polyethylenes.

Authors:  Carlos Roberto Schwartsmann; Leandro de Freitas Spinelli; Leonardo Carbonera Boschin; Ramiro Zilles Gonçalves; Anthony Kerbes Yépez; Telmo Roberto Strohaecker; Ralf Wellis de Souza
Journal:  Rev Bras Ortop       Date:  2013-12-19

8.  Second-generation uncemented total hip arthroplasty: a minimum 20-year follow-up.

Authors:  Jeffrey R McLaughlin; Kyla R Lee; Mary Ann Johnson
Journal:  Bone Jt Open       Date:  2021-01-13

9.  Long-Term results of total Hip Arthroplasty performed using a cementless expansive Acetabular Cup and Spotorno Femoral Stem.

Authors:  Engin Carkci; Ayse Esin Polat; Yusuf Ozturkmen; Tolga Tuzuner
Journal:  Pak J Med Sci       Date:  2021 Jan-Feb       Impact factor: 1.088

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.