Literature DB >> 21762991

First eye prediction error improves second eye refractive outcome results in 2129 patients after bilateral sequential cataract surgery.

Petros Aristodemou1, Nathaniel E Knox Cartwright, John M Sparrow, Robert L Johnston.   

Abstract

OBJECTIVE: To define theoretical correction factors for second-eye intraocular lens (IOL) power adjustment based on first eye refractive prediction error (PE).
DESIGN: Database study. PARTICIPANTS: We included 2129 patients who underwent bilateral sequential phacoemulsification cataract surgery with the same IOL model.
METHODS: Retrospectively calculated PEs (Hoffer Q, Holladay 1, SRK/T) were analyzed for association between paired eyes, examining the effect of interocular differences in axial length (AL) and corneal power. A range of correction factors (CF) derived from the first eye PE were applied to the second eye PE using optimized and non-optimized IOL constants (IOLCs). MAIN OUTCOME MEASURES: Second eye mean absolute error (MAE).
RESULTS: Prediction errors of paired eyes were correlated. Interocular corneal power differences exceeding 0.60 diopters (D) were associated with a weaker correlation but interocular AL differences did not affect the correlation. When a 50% CF was applied to second eyes of patients with a first eye PE between ±0.50 and ±1.50 D, it improved refractive outcomes from 30%, 56%, and 92% to 42%, 75%, and 96% within ±0.25 D, ±0.50 D and ±1.00 D, respectively, and reduced the MAE from 0.49 to 0.37 D (P<0.0001). For first eye PE below ±0.50 D, a 50% CF reduced the MAE from 0.32 to 0.30 D (P<0.00001). A 50% CF also reduces second eye MAE for eyes with nonoptimized IOLCs.
CONCLUSIONS: A 50% CF reduces second eye PE when either first eye optimized PE is within ±1.50 D or when nonoptimized IOLCs are used. The correlation is weaker when interocular corneal power differences are >0.60 D.
Copyright © 2011 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21762991     DOI: 10.1016/j.ophtha.2011.05.010

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  13 in total

1.  Cataract update.

Authors:  Ali A Al-Rajhi
Journal:  Saudi J Ophthalmol       Date:  2012-01

Review 2.  Optimising biometry for best outcomes in cataract surgery.

Authors:  R Sheard
Journal:  Eye (Lond)       Date:  2013-12-06       Impact factor: 3.775

Review 3.  Recurring themes during cataract assessment and surgery.

Authors:  Shruti Chandra; Sobha Sivaprasad; Paul G Ursell; Khayam Naderi; David O'Brart; Amar Alwitry; Mayank A Nanavaty
Journal:  Eye (Lond)       Date:  2021-04-29       Impact factor: 4.456

4.  Influence of the prediction error of the first eye undergoing cataract surgery on the refractive outcome of the fellow eye.

Authors:  Ludmilla Gorodezky; Babac Ae Mazinani; Niklas Plange; Peter Walter; Martin Wenzel; Gernot Roessler
Journal:  Clin Ophthalmol       Date:  2014-10-31

5.  Immediate sequential bilateral cataract surgery: A 5-year retrospective analysis of 2470 eyes from a tertiary care eye center in South India.

Authors:  Sri Ganesh; Sheetal Brar; Rohit Sreenath
Journal:  Indian J Ophthalmol       Date:  2017-05       Impact factor: 1.848

6.  Correlation of binocular refractive error and calculation of intraocular Lens power for the second eye.

Authors:  Pengcheng Zhang; Yuhuan Yang; Hong Yan; Jie Zhang; Weijia Yan
Journal:  BMC Ophthalmol       Date:  2020-06-18       Impact factor: 2.209

7.  Safety of Simultaneous Bilateral Intraocular Surgery Under General Anesthesia in Pediatric Patients.

Authors:  Elham R Al-Tamimi
Journal:  Saudi J Med Med Sci       Date:  2016-08-11

8.  Interocular biometric parameters comparison measured with swept-source technology.

Authors:  César Albarrán-Diego; Francisco Poyales; Esther López-Artero; Nuria Garzón; María García-Montero
Journal:  Int Ophthalmol       Date:  2021-08-21       Impact factor: 2.031

9.  Postoperative refraction in the second eye having cataract surgery.

Authors:  Christopher T Leffler; Martin Wilkes; Juliana Reeves; Muneera A Mahmood
Journal:  ISRN Ophthalmol       Date:  2011-12-10

10.  Disruptive Innovation and Refractive IOLs: How the Game Will Change With Adjustable IOLs.

Authors:  David F Chang
Journal:  Asia Pac J Ophthalmol (Phila)       Date:  2019 Nov-Dec
View more