Computer-aided detection scheme for sentinel lymph nodes in lymphoscintigrams using symmetrical property around mapped injection point.

Ryohei Nakayama1, Akiyoshi Hizukuri, Koji Yamamoto, Nobuo Nakako, Naoki Nagasawa, Kan Takeda.   

Abstract

It is difficult to detect sentinel lymph nodes (SLNs) around an injection point of radiopharmaceuticals mapped in lymphoscintigrams. The purpose of this study was to develop a computer-aided detection (CAD) scheme for SLNs by a subtraction technique using the symmetrical property in the mapped injection point. Our database consisted of 78 lymphoscintigrams with 86 SLNs. In our CAD scheme, the mapped injection point of radiopharmaceuticals was first segmented from the lymphoscintigram using a gray-level thresholding technique. Lymphoscintigram was then divided into four regions by vertical and horizontal straight lines through the center of the segmented injection point. One of the four divided regions was defined as the target region. The correlation coefficients based on pixel values were calculated between the target region and each of the other three regions. The region with the highest correlation coefficient among three regions was selected as the similar region to the target region. The values of pixels on the target region were subtracted by the values of the corresponding pixels on the similar region. This procedure was repeated until every divided region had been used as target region. SLNs were segmented by applying a gray-level thresholding technique to the subtracted image. With our CAD scheme, sensitivity and the number of false positives were 95.3% (82/86) and 2.51 per image, respectively. Our CAD scheme achieved a high level of detection accuracy, and would have a great potential in assisting physicians to detect SLNs in lymphoscintigrams.

Mesh:

Year:  2012        PMID: 21725620      PMCID: PMC3264709          DOI: 10.1007/s10278-011-9396-8

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  22 in total

Review 1.  Radiopharmaceuticals for lymphoscintigraphy: including dosimetry and radiation considerations.

Authors:  D Eshima; T Fauconnier; L Eshima; J R Thornback
Journal:  Semin Nucl Med       Date:  2000-01       Impact factor: 4.446

2.  [Contrast enhancement for sentinel lymph node imaging].

Authors:  K Ogawa; H Fujii; Y Kitagawa; A Kubo
Journal:  Kaku Igaku       Date:  2001-07

3.  Improved contralateral subtraction images by use of elastic matching technique.

Authors:  Q Li; S Katsuragawa; K Doi
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

4.  Local contralateral subtraction based on bilateral symmetry of lung for reduction of false positives in computerized detection of pulmonary nodules.

Authors:  Hiroyuki Yoshida
Journal:  IEEE Trans Biomed Eng       Date:  2004-05       Impact factor: 4.538

5.  Computerized detection of masses in digital mammograms: analysis of bilateral subtraction images.

Authors:  F F Yin; M L Giger; K Doi; C E Metz; C J Vyborny; R A Schmidt
Journal:  Med Phys       Date:  1991 Sep-Oct       Impact factor: 4.071

6.  Some practical issues of experimental design and data analysis in radiological ROC studies.

Authors:  C E Metz
Journal:  Invest Radiol       Date:  1989-03       Impact factor: 6.016

7.  Maximum likelihood analysis of free-response receiver operating characteristic (FROC) data.

Authors:  D P Chakraborty
Journal:  Med Phys       Date:  1989 Jul-Aug       Impact factor: 4.071

8.  Technetium-99m sulfide colloid preparation for scintigraphy of the reticuloendothelial system.

Authors:  R Bertil; R Persson; Y Naversten
Journal:  Acta Radiol Ther Phys Biol       Date:  1970-12

9.  Relation of number of positive axillary nodes to the prognosis of patients with primary breast cancer. An NSABP update.

Authors:  B Fisher; M Bauer; D L Wickerham; C K Redmond; E R Fisher; A B Cruz; R Foster; B Gardner; H Lerner; R Margolese
Journal:  Cancer       Date:  1983-11-01       Impact factor: 6.860

10.  Validation and statistical power comparison of methods for analyzing free-response observer performance studies.

Authors:  Dev P Chakraborty
Journal:  Acad Radiol       Date:  2008-12       Impact factor: 3.173

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.