Literature DB >> 21723802

The EUROP total knee prosthesis: a ten-year follow-up study of a posterior cruciate-retaining design.

A Mouttet1, M-L Louis, V Sourdet.   

Abstract

INTRODUCTION: The success of total knee arthroplasty is measured by pain relief, functional recovery, and implant survival duration. The aim of the present study was to evaluate the long-term clinical, functional and radiological results of the posterior cruciate ligament (PCL)-retaining fixed bearing EUROP implant. HYPOTHESIS: The long-term results of EUROP implants are similar to those reported with comparable prostheses. PATIENTS AND METHODS: We performed a prospective, monocentric study of a series of 121 cemented EUROP total knee arthroplasties, implanted between 1994 and 1996 in 117 patients mean age 73. A clinical and radiological evaluation was performed at 10 years of follow-up according to the International Knee Society (IKS) score. Twenty-three patients died, 14 were lost to follow-up, 43 underwent clinical and radiological evaluation and 37 were questioned by telephone.
RESULTS: The preoperative IKS knee score was 31 points (0-60) and increased to 88 points (30-98) at final follow-up, IKS function increased from 40 (0-90) to 80 points (25-100). Radiolucencies were observed in 56% of the condyles and 60% of tibial plates. Ninety-three percent of these radiolucent lines were less than 1mm wide. Three patients underwent revision TKA at 32 months, eight and 11 years respectively. Global implant survival was 99% at five years, 97.8% at 10 years and 95.8% at 12 years. DISCUSSION: The clinical and radiological results of the cruciate-retaining fixed bearing EUROP total knee arthroplasties, with three cases of revision arthroplasty at 12 years of follow-up are satisfactory and comparable to similar implants.
Copyright © 2011 Elsevier Masson SAS. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21723802     DOI: 10.1016/j.otsr.2011.02.013

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  1 in total

1.  The self-aligning knee prosthesis: clinical and radiological outcome and survival analysis of a cruciate retaining meniscal bearing knee at 10-year follow-up.

Authors:  R A van Stralen; P G Anderson; A B Wymenga
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2013-07-10       Impact factor: 4.342

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.