Literature DB >> 21714424

Clinical and hemodynamic outcomes after aortic valve replacement with stented and stentless pericardial xenografts: a propensity-matched analysis.

Augusto D'Onofrio1, Giovanni Domenico Cresce, Irene Bolgan, Paolo Magagna, Caterina Piccin, Stefano Auriemma, Alessandro Fabbri.   

Abstract

BACKGROUND AND AIM OF THE STUDY: Pericardial aortic xenografts have demonstrated excellent durability, and also freedom from tissue failure and from endocarditis. The aim of this single-center propensity-matched study was to compare the clinical and hemodynamic results of aortic valve replacement (AVR) with that for stented and stentless pericardial bioprostheses.
METHODS: A total of 111 consecutive AVRs with the Perimount Magna stented valve, performed between December 2002 and December 2007, and 150 consecutive AVRs with the Pericarbon Freedom stentless bioprosthesis, performed between July 1999 and December 2007, was reviewed. Based on a propensity-score analysis, 81 matched pairs were created. The mean age of the 162 patients was 74 +/- 7 years. In total, 31 (38%) and 39 (48%) associated procedures were performed in the stented and stentless groups, respectively (p = 0.20). The mean follow up for stented-valve patients was 20 +/- 9 months (range: 6-35 months), and for stentless-valve patients was 36 +/- 21 months (range: 6-83 months) (p < 0.05).
RESULTS: The operative mortality was 4.9% (n = 4) and 6.2% (n = 5) (p = 0.77) in the stented- and stentless-valve groups, respectively. The cumulative actuarial freedom from structural valve deterioration, non-structural dysfunction, reoperation, prosthetic valve endocarditis, hemorrhage and thromboembolic events after 24 months were 98.4 +/- 1.5% and 96.4 +/- 2.6% (p = 0.12) for the stented- and stentless-valve groups, respectively; the actuarial survival rates after 24 months were 82.6 +/- 5.2% and 83.8 +/- 4.3% (p = 0.82), respectively, while freedom from valve-related death was 98.4 +/- 1.5% and 97.2 +/- 1.9% (p = 0.56), respectively. No statistically significant differences were identified between the peak and mean transaortic gradients.
CONCLUSION: Both, the stented and stentless pericardial aortic xenografts provided good clinical and hemodynamic results. However, the study results failed to demonstrate any difference regarding early outcomes between the two groups.

Entities:  

Mesh:

Year:  2011        PMID: 21714424

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  3 in total

1.  One-step strategy for cartilage repair using acellular bone matrix scaffold based in situ tissue engineering technique in a preclinical minipig model.

Authors:  Linghui Dai; Zhenming He; Yanfang Jiang; Xin Zhang; Shuang Ren; Jingxian Zhu; Zhenxing Shao; Hongjie Huang; Jiying Zhang; Xin Fu; Xiaoning Duan; Xiaoqing Hu; Yingfang Ao
Journal:  Am J Transl Res       Date:  2019-10-15       Impact factor: 4.060

2.  Multipotent Stromal Cells Outperform Chondrocytes on Cartilage-Derived Matrix Scaffolds.

Authors:  K E M Benders; W Boot; S M Cokelaere; P R Van Weeren; D Gawlitta; H J Bergman; D B F Saris; W J A Dhert; J Malda
Journal:  Cartilage       Date:  2014-10       Impact factor: 4.634

3.  Early and mid-term haemodynamic performance and clinical outcomes of St. Jude Medical Trifecta™ valve.

Authors:  Renata Raimundo; Soraia Moreira; Francisca Saraiva; Rui J Cerqueira; Pedro Teixeira; Elson Salgueiro; André Lourenço; Mário J Amorim; Jorge Almeida; Paulo Pinho; Adelino F Leite-Moreira
Journal:  J Thorac Dis       Date:  2018-02       Impact factor: 2.895

  3 in total