| Literature DB >> 21693792 |
Abstract
Tomographic systems employing truncated projections have been developed for parallel and fan beam collimation and for cone beam CT but the idea has not been extensively explored in pinhole single photon emission computed tomography (SPECT). In this paper, we explore the sampling requirements and system performance of SPECT systems with asymmetric pinhole collimators and truncated projections. We demonstrate that complete 3D sampling can be achieved by using multiple detectors with truncated asymmetric pinholes, offset axially from each other, and a spiral orbit. The use of truncated projections can be exploited in the design of pinhole SPECT systems by moving the pinholes closer to the subject, resulting in increased sensitivity and improved spatial resolution. Truncated and untruncated pinhole systems were evaluated using the contrast-to-noise ratio (CNR) calculated from the linearized local impulse response as a figure of merit. The CNR for the truncated pinhole system was up to 60% greater than that for the untruncated system at matched resolution for a source voxel near the centre of a uniform phantom and 30% greater at the edge. We conclude that an object can be reconstructed from asymmetric pinholes with truncated projections, which leads to potentially important design considerations and applications in single- and multi-pinhole SPECT.Mesh:
Year: 2011 PMID: 21693792 DOI: 10.1088/0031-9155/56/13/023
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609