The silicon-photomultiplier (Si-PM) is a promising photodetector, especially for integrated PET/MRI systems, due to its small size, high gain, and low sensitivity to static magnetic fields. The major problem using a Si-PM-based PET system within the MRI system is the interference between the PET and MRI units. We measured the interference by combining a Si-PM-based PET system with a permanent-magnet MRI system. When the RF signal-induced pulse height exceeded the lower energy threshold level of the PET system, interference between the Si-PM-based PET system and MRI system was detected. The prompt as well as the delayed coincidence count rates of the Si-PM-based PET system increased significantly. These noise counts produced severe artifacts on the reconstructed images of the Si-PM-based PET system. In terms of the effect of the Si-PM-based PET system on the MRI system, although no susceptibility artifact was observed on the MR images, electronic noise from the PET detector ring was detected by the RF coil and reduced the signal-to-noise ratio (S/N) of the MR images. The S/N degradation of the MR images was reduced when the distance between the RF coil and the Si-PM-based PET system was increased. We conclude that reducing the interference between the PET and MRI systems is essential for achieving the optimum performance of integrated Si-PM PET/MRI systems.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21693791     DOI: 10.1088/0031-9155/56/13/026

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Development of 1.45-mm resolution four-layer DOI-PET detector for simultaneous measurement in 3T MRI.

Authors:  Fumihiko Nishikido; Atsushi Tachibana; Takayuki Obata; Naoko Inadama; Eiji Yoshida; Mikio Suga; Hideo Murayama; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2014-10-28

2.  After-pulsing, cross-talk, dark-count, and gain of MPPC under 7-T static magnetic field.

Authors:  Yoshiyuki Hirano; Fumihiko Nishikido; Daisuke Kokuryo; Taiga Yamaya
Journal:  Radiol Phys Technol       Date:  2016-05-17

3.  New shielding configurations for a simultaneous PET/MRI scanner at 7T.

Authors:  Bo J Peng; Yibao Wu; Simon R Cherry; Jeffrey H Walton
Journal:  J Magn Reson       Date:  2013-11-14       Impact factor: 2.229

4.  Low eddy current RF shielding enclosure designs for 3T MR applications.

Authors:  Brian J Lee; Ronald D Watkins; Chen-Ming Chang; Craig S Levin
Journal:  Magn Reson Med       Date:  2017-06-06       Impact factor: 4.668

Review 5.  Principles of Simultaneous PET/MR Imaging.

Authors:  Ciprian Catana
Journal:  Magn Reson Imaging Clin N Am       Date:  2017-05       Impact factor: 2.266

6.  Simulating Silicon Photomultiplier Response to Scintillation Light.

Authors:  Abhinav K Jha; Herman T van Dam; Matthew A Kupinski; Eric Clarkson
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

7.  Image quality assessment of automatic three-segment MR attenuation correction vs. CT attenuation correction.

Authors:  Sasan Partovi; Andres Kohan; Chiara Gaeta; Christian Rubbert; Jose L Vercher-Conejero; Robert S Jones; James K O'Donnell; Patrick Wojtylak; Peter Faulhaber
Journal:  Am J Nucl Med Mol Imaging       Date:  2013-04-09

8.  PET/MRI insert using digital SiPMs: Investigation of MR-compatibility.

Authors:  Jakob Wehner; Bjoern Weissler; Peter Dueppenbecker; Pierre Gebhardt; David Schug; Walter Ruetten; Fabian Kiessling; Volkmar Schulz
Journal:  Nucl Instrum Methods Phys Res A       Date:  2014-01-11       Impact factor: 1.455

Review 9.  MRI/PET Imaging in elevated PSA and localized prostate cancer: a narrative review.

Authors:  Subodh K Regmi; Niranjan Sathianathen; Thomas E Stout; Badrinath R Konety
Journal:  Transl Androl Urol       Date:  2021-07

10.  FPGA-based RF interference reduction techniques for simultaneous PET-MRI.

Authors:  P Gebhardt; J Wehner; B Weissler; R Botnar; P K Marsden; V Schulz
Journal:  Phys Med Biol       Date:  2016-04-06       Impact factor: 3.609

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.