Richard D Powell1, William C Whitworth, John Bernardo, Patrick K Moonan, Gerald H Mazurek.
Abstract
INTRODUCTION: Interferon gamma (IFN-γ) release assays, such as QuantiFERON®-TB Gold test (QFT-G) and QuantiFERON®-TB Gold In-Tube test (QFT-GIT) are designed to detect M. tuberculosis (Mtb) infection. Recognition of unusual IFN-γ measurements may help indicate inaccurate results.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21687702 PMCID: PMC3110578 DOI: 10.1371/journal.pone.0020061
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Interpretation Criteria used for QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tubea Tests.
| Interpretation | [Nil] | TB Response | Mitogen Response |
| Positive | Any | ≥0.35 IU/mL and ≥50% of [Nil] | Any |
| Negative | ≤0.7 IU/mL | <0.35 IU/mL | ≥0.5 IU/mL |
| Indeterminate | ≤0.7 IU/mL | <0.35 IU/mL | <0.5 IU/mL |
| >0.7 IU/mL | <50% of Nil | Any |
Interpretation criteria used for QFT-GIT differed from that approved by the FDA in Nov, 2008.
“[Nil]” is the IFN-γ concentration in plasma from unstimulated blood.
“TB Response” for the QuantiFERON-TB Gold test is the higher IFN-γ concentration in plasma from blood stimulated by a cocktail of peptides representing ESAT-6 or CFP10, minus [Nil].
“TB Response” for the QuantiFERON-TB Gold In-Tube test is the IFN-γ concentration in plasma from blood stimulated by a single cocktail of peptides representing ESAT-6, CFP10, and part of TB7.7 minus [Nil].
“Mitogen Response” is the IFN-γ concentration in plasma from mitogen stimulated blood minus [Nil].
Frequency of Unusual IFN-γ Measurements with QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube Testing.
| QFT-G | QFT-GIT | Total | OR (95% CI) |
| |
| # of tests | 2741 | 2568 | 5309 | ||
| # of subjects | 1696 | 1653 | 1728 | ||
| # (%) of tests with | 136 (5.0) | 98 (3.8) | 234 (4.4) | 1.32 (1.01–1.72) | 0.042 |
| # (%) of subjects with | 122 | 83 | 162 | ||
| # (%) of tests with | 41 (1.5) | 67 (2.6) | 108 (2.0) | 0.57 (0. 38–0.84) | 0.004 |
| # (%) of subjects with | 38 | 59 | 85 | ||
| # (%) of tests with | 18 (0.7) | 4 (0.2) | 22 (0.4) | 4.24 (1.42–12.54) | 0.005 |
| # (%) of subjects with | 17 | 4 | 21 | ||
| # (%) of tests with | 27 (1.0) | 14 (0.5) | 41 (0.8) | 1.82 (0.95–3.47) | 0.067 |
| # (%) of subjects with | 27 | 14 | 39 | ||
| # (%) of tests with any 1 or more Unusual IFN-γ Measurements | 171 (6.2) | 164 (6.4) | 335 (6.3) | ||
| # (%) of subjects with any 1 or more Unusual IFN-γ Measurements | 154 | 140 | 236 |
“QFT-G” is QuantiFERON-TB Gold test.
“QFT-GIT” is QuantiFERON-TB Gold In-Tube test.
“Total # of subjects” with unusual measurement may be less than the sum of the “# of subjects” with unusual QFT-G and QFT-GIT measurements because subjects may have unusual measurements with both tests.
“OR (95% CI)” is odds ratio and 95% confidence interval for an unusual IFN-γ measurement in QFT-G compared to referent, QFT-GIT.
“HNC” is High Nil Concentration (i.e. [Nil] over 0.7 IU/mL).
“LMR” is Low Mitogen Response (i.e. Mitogen Response under 0.5 IU/mL).
“VLMR” is Very Low Mitogen Response (i.e. Mitogen Response ≤−0.5 IU/mL).
“VLAR” is Very Low Antigen Response (i.e. response to a Mtb antigen ≤−0.35 IU/mL and ≤−0.5 times [Nil]).
Figure 1Number of subjects with unusual IFN-γ measurements.
The number of subjects with unusual IFN-γ measurements categorized as High Nil Concentration (HNC), Low Mitogen Response (LMR), Very Low Mitogen Response (VLMR), and Very Low Antigen Response (VLAR) that occurred once (dark gray bars) or more than once (light gray bars) among 1,725 subjects who had 5,309 QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube Tests performed.
Subjects in Each Group with Unusual IFN-γ Measurements.
| Group | ≥1 HNC | ≥1 LMR | ≥1 VLMR | ≥1 VLAR | |
| Total | n (%) | n (%) | n (%) | n (%) | |
| Employee Screening | 864 | 35 (4.1) | 12 (1.4) | 0 (0) | 12 (1.4) |
| Contacts | 683 | 103 (15.1) | 50 (7.3) | 11 (1.6) | 23 (3.4) |
| TB Suspect | 181 | 24(13.3) | 23 (12.7) | 10 (5.5) | 4 (2.2) |
|
| <0.001 | <0.001 | <0.001 | 0.03 |
“HNC” is High Nil Concentration (i.e. [Nil] over 0.7 IU/mL).
“LMR” is Low Mitogen Response (i.e. Mitogen Response under 0.5 IU/mL).
“VLMR” is Very Low Mitogen Response (i.e. Mitogen Response ≤−0.5 IU/mL).
“VLAR” is Very Low Antigen Response (i.e. response to a Mtb antigen ≤−0.35 IU/mL and ≤−0.5 IU/mL).
Figure 2Number of subjects with unusual IFN-γ measurements when QuantiFERON-TB Gold tests were repeated.
The number of subjects with unusual IFN-γ measurements categorized as High Nil Concentration (HNC), Low Mitogen Response (LMR), Very Low Mitogen Response (VLMR), and Very Low Antigen Response (VLAR) that occurred only with the initial test (light gray bars), with the repeat test (dark gray bars), or in both the initial and repeat test (medium gray bars) among 1,044 Subjects with Initial and Repeat QuantiFERON-TB Gold Test Results.
Figure 3Number of subjects with unusual IFN-γ measurements when QuantiFERON-TB Gold In-Tube tests were repeated.
The number of subjects with unusual IFN-γ measurements categorized as High Nil Concentration (HNC), Low Mitogen Response (LMR), Very Low Mitogen Response (VLMR), and Very Low Antigen Response (VLAR) that occurred only with the initial test (light gray bars), with the repeat test (dark gray bars), or in both the initial and repeat test (medium gray bars) among 915 subjects with initial and repeat QuantiFERON-TB Gold In-Tube test results.
Figure 4Number of subjects with unusual IFN-γ measurements when QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube tests were run simultaneously.
The number of subjects with unusual IFN-γ measurements categorized as High Nil Concentration (HNC), Low Mitogen Response (LMR), Very Low Mitogen Response (VLMR), and Very Low Antigen Response (VLAR) that occurred only with the QuantiFERON-TB Gold test (light gray bars), with the QuantiFERON-TB Gold In-Tube test (dark gray bars), or in both the QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube test (medium gray bars) among 1,617 subjects with simultaneously performed QuantiFERON-TB Gold and QuantiFERON-TB Gold In-Tube Test.