Literature DB >> 21680862

The temperature and humidity in a low-flow anesthesia workstation with and without a heat and moisture exchanger.

Jair de Castro1, Fernanda Bolfi, Lidia R de Carvalho, Jose R C Braz.   

Abstract

BACKGROUND: The Dräger Primus anesthesia workstation has a built-in hotplate to heat the patient's exhaled gas. The fresh gas flow is mixed with the heated exhaled gas as they pass through the soda lime canister. A heat and moisture exchanger (HME) may also be used to further heat and humidify the inhaled gas. In this study we measured the temperature and humidity of the inhaled gas coming from the Dräger Primus with or without a HME.
METHODS: Thirty female patients were randomly divided into 2 groups and their lungs ventilated by the Primus Dräger anesthesia workstation with or without a HME. The humidity and temperature of the inhaled gas were measured 15, 30, 60, 90, and 120 minutes after connecting the patient to the breathing circuit.
RESULTS: After 120 minutes of ventilation with a low-flow breathing circuit, the temperatures of inhaled gas were 25°C ± 1°C and 30°C ± 2°C without and with HME, respectively, with a statistically significant difference between groups (P < 0.001) with 95% confidence interval (CI) of 3.80°C to 6.40°C; and the absolute humidity values of the inhaled gas were 20.5 ± 3.6 mgH(2)O · L(-1) and 30 ± 2 mgH(2)O · L(-1) without and with HME, respectively, with a statistically significant difference between groups (P < 0.001) with 95% CI of 7.37°C to 13.03°C.
CONCLUSIONS: The Primus anesthesia workstation partially humidifies the inspired gas when a low fresh gas flow is used. Insertion of an HME increases the humidity in inhaled gas, bringing it close to physiological values.

Entities:  

Mesh:

Year:  2011        PMID: 21680862     DOI: 10.1213/ANE.0b013e31822402df

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  9 in total

1.  Effect of a new heated and humidified breathing circuit with a fluid-warming device on intraoperative core temperature: a prospective randomized study.

Authors:  Eugene Kim; Sue-Young Lee; Young-Jin Lim; Jung-Yoon Choi; Young-Tae Jeon; Jung-Won Hwang; Hee-Pyoung Park
Journal:  J Anesth       Date:  2015-03-14       Impact factor: 2.078

2.  Inadvertent hypothermia and acute kidney injury (AKI) in neonates undergoing gastrointestinal surgeries: a retrospective study.

Authors:  Yu Cui; Rong Cao; Lingmei Deng
Journal:  J Perinatol       Date:  2021-08-19       Impact factor: 2.521

3.  Comparison of the temperature and humidity in the anesthetic breathing circuit among different anesthetic workstations: Updated guidelines for reporting parallel group randomized trials.

Authors:  Yoon Ji Choi; Sam Hong Min; Jeong Jun Park; Jang Eun Cho; Seung Zhoo Yoon; Suk Min Yoon
Journal:  Medicine (Baltimore)       Date:  2017-06       Impact factor: 1.889

4.  The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial.

Authors:  Sergius A R de Oliveira; Lorena M C Lucio; Norma S P Modolo; Yoko Hayashi; Mariana G Braz; Lídia R de Carvalho; Leandro G Braz; José Reinaldo C Braz
Journal:  PLoS One       Date:  2017-01-27       Impact factor: 3.240

5.  A New Method for In Vivo Analysis of the Performances of a Heat and Moisture Exchanger (HME) in Mechanically Ventilated Patients.

Authors:  Matteo Filippini; Mauro Serpelloni; Valeria Quaranta; Paolo Bellitti; Emilio Sardini; Nicola Latronico
Journal:  Pulm Med       Date:  2019-02-26

6.  The low fresh gas flow anesthesia and hypothermia in neonates undergoing digestive surgeries: a retrospective before-after study.

Authors:  Yu Cui; Yu Wang; Rong Cao; Gen Li; Lingmei Deng; Jia Li
Journal:  BMC Anesthesiol       Date:  2020-09-03       Impact factor: 2.217

7.  Effects of different fresh gas flows and different anesthetics on airway temperature and humidity in surgical patients: a prospective observational study.

Authors:  Irem Gumus Ozcan; Ozkan Onal; Aysun Ozdemirkan; Ali Saltali; Mehmet Sari
Journal:  Med Gas Res       Date:  2022 Jul-Sep

8.  Impact of the Hormonal Status in Women on Intraoperative Hypothermia during Laparoscopic Gynecologic Surgery when Considering the Fresh Gas Flow Rate: A Retrospective Study.

Authors:  Cheol Lee; Juhwan Lee; Gongheui Lee; SeongNam Park; Myeongjong Lee; Hyungtae Kim
Journal:  Biomed Res Int       Date:  2022-02-08       Impact factor: 3.411

9.  Huang Qin Hua Shi decoction for high-temperature- and high-humidity-induced cognitive-behavioral disorder in rats is associated with deactivation of the hypothalamic-pituitary-adrenal axis.

Authors:  Yong Luo; Min Yang; Mingyang Guo; Xiaolong Zhong; Yonghe Hu
Journal:  J Int Med Res       Date:  2019-09-23       Impact factor: 1.671

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.