Flow dynamics of a novel counterpulsation device characterized by CFD and PIV modeling.

G A Giridharan1, C Lederer, A Berthe, L Goubergrits, J Hutzenlaub, M S Slaughter, R D Dowling, P A Spence, S C Koenig.   

Abstract

BACKGROUND: Historically, single port valveless pneumatic blood pumps have had a high incidence of thrombus formation due to areas of blood stagnation and hemolysis due to areas of high shear stress.
METHODS: To ensure minimal hemolysis and favorable blood washing characteristics, particle image velocimetry (PIV) and computational fluid dynamics (CFD) were used to evaluate the design of a new single port, valveless counterpulsation device (Symphony). The Symphony design was tested in 6-h acute (n=8), 5-day (n=8) and 30-day (n=2) chronic experiments in a calf model (Jersey, 76 kg). Venous blood samples were collected during acute (hourly) and chronic (weekly) time courses to analyze for temporal changes in biochemical markers and quantify plasma free hemoglobin. At the end of the study, animals were euthanized and the Symphony and end-organs (brain, liver, kidney, lungs, heart, and spleen) were examined for thrombus formations.
RESULTS: Both the PIV and the CFD showed the development of a strong moving vortex during filling phase and that blood exited the Symphony uniformly from all areas during ejection phase. The laminar shear stresses estimated by CFD remained well below the hemolysis threshold of 400 Pa inside the Symphony throughout filling and ejection phases. No areas of persistent blood stagnation or flow separation were observed. The maximum plasma free hemoglobin (<10mg/dl), average platelet count (pre-implant = 473 ± 56 K/μl and post-implant = 331 ± 62 K/μl), and average hematocrit (pre-implant = 31 ± 2% and post-implant = 29 ± 2%) were normal at all measured time-points for each test animal in acute and chronic experiments. There were no changes in measures of hepatic function (ALP, ALT) or renal function (creatinine) from pre-Symphony implantation values. The necropsy examination showed no signs of thrombus formation in the Symphony or end organs.
CONCLUSIONS: These data suggest that the designed Symphony has good washing characteristics without persistent areas of blood stagnation sites during the entire pump cycle, and has a low risk of hemolysis and thrombus formations.
Copyright © 2011 IPEM. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21680224      PMCID: PMC5898629          DOI: 10.1016/j.medengphy.2011.05.008

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  27 in total

Review 1.  Intra-aortic balloon counterpulsation.

Authors:  Jeffrey C Trost; L David Hillis
Journal:  Am J Cardiol       Date:  2006-03-20       Impact factor: 2.778

Review 2.  LVAD-induced reverse remodeling: basic and clinical implications for myocardial recovery.

Authors:  Daniel Burkhoff; Stefan Klotz; Donna M Mancini
Journal:  J Card Fail       Date:  2006-04       Impact factor: 5.712

3.  A helical PTFE arteriovenous access graft to swirl flow across the distal anastomosis: results of a preliminary clinical study.

Authors:  H J T A M Huijbregts; P J Blankestijn; C G Caro; N J W Cheshire; M T C Hoedt; R P Tutein Nolthenius; F L Moll
Journal:  Eur J Vasc Endovasc Surg       Date:  2006-12-11       Impact factor: 7.069

4.  Artificial heart research and present status of clinical application in Japan.

Authors:  Kou Imachi
Journal:  ASAIO J       Date:  2006 Jan-Feb       Impact factor: 2.872

5.  Hemodynamic and left ventricular pressure-volume responses to counterpulsation in mock circulation and acute large animal models.

Authors:  S C Koenig; G M Pantalos; K N Litwak; K J Gillars; G A Giridharan; M Maguire; P A Spence
Journal:  Conf Proc IEEE Eng Med Biol Soc       Date:  2004

6.  Multimodal flow visualization and optimization of pneumatic blood pump for sorbent hemodialysis system.

Authors:  Fangjun Shu; Robert Parks; John Maholtz; Steven Ash; James F Antaki
Journal:  Artif Organs       Date:  2009-04       Impact factor: 3.094

7.  Hemolytic differences among artificial cardiac valves used in a ventricular assist pump.

Authors:  G G Billy; C A Miller; M N Pallone; J H Donachy; W S Pierce
Journal:  Artif Organs       Date:  1995-04       Impact factor: 3.094

8.  And hemolysis goes on: ventricular assist device in combination with veno-venous hemofiltration.

Authors:  Heyman Luckraz; Michael Woods; Stephen R Large
Journal:  Ann Thorac Surg       Date:  2002-02       Impact factor: 4.330

Review 9.  Mechanical device-based methods of managing and treating heart failure.

Authors:  Donna Mancini; Daniel Burkhoff
Journal:  Circulation       Date:  2005-07-19       Impact factor: 29.690

Review 10.  The role of bridge to transplantation: should LVAD patients be transplanted?

Authors:  Emma J Birks; Magdi H Yacoub; Nicholas R Banner; Asghar Khaghani
Journal:  Curr Opin Cardiol       Date:  2004-03       Impact factor: 2.161

View more
  5 in total

1.  Efficacy of Subcutaneous Electrocardiogram Leads for Synchronous Timing During Chronic Counterpulsation Therapy.

Authors:  Stephen R Carnahan; Steven C Koenig; Michael A Sobieski; Erin M Schumer; Gretel Monreal; Yu Wang; Young Choi; Brek J Meuris; Landon H Tompkins; Zhongjun J Wu; Mark S Slaughter; Guruprasad A Giridharan
Journal:  ASAIO J       Date:  2017 Mar/Apr       Impact factor: 2.872

Review 2.  Miniaturization of mechanical circulatory support systems.

Authors:  Guruprasad A Giridharan; Thomas J Lee; Mickey Ising; Michael A Sobieski; Steven C Koenig; Laman A Gray; Mark S Slaughter
Journal:  Artif Organs       Date:  2012-08       Impact factor: 3.094

Review 3.  Device-based treatment options for heart failure with preserved ejection fraction.

Authors:  Chihiro Miyagi; Takuma Miyamoto; Jamshid H Karimov; Randall C Starling; Kiyotaka Fukamachi
Journal:  Heart Fail Rev       Date:  2021-01-12       Impact factor: 4.214

Review 4.  Evidence of clinical efficacy of counterpulsation therapy methods.

Authors:  M Capoccia; C T Bowles; J R Pepper; N R Banner; A R Simon
Journal:  Heart Fail Rev       Date:  2015-05       Impact factor: 4.214

5.  A Valveless Pulsatile Pump for Heart Failure with Preserved Ejection Fraction: Hemo- and Fluid Dynamic Feasibility.

Authors:  Andreas Escher; Young Choi; Fraser Callaghan; Bente Thamsen; Ulrich Kertzscher; Martin Schweiger; Michael Hübler; Marcus Granegger
Journal:  Ann Biomed Eng       Date:  2020-03-30       Impact factor: 3.934

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.