Daniel C Paech1, Adèle R Weston.
Abstract
BACKGROUND: This systematic review summarized recent evidence pertaining to the clinical effectiveness of 64-slice or higher computed tomography angiography (CTA) in patients with suspected coronary artery disease (CAD). If CTA proves to be a successful diagnostic performance measure, it could prevent the use of invasive diagnostic procedures in some patients. This would provide multiple health and cost benefits, particularly for under resourced areas where invasive coronary angiography is not always available.Entities:
Mesh:
Year: 2011 PMID: 21679468 PMCID: PMC3141758 DOI: 10.1186/1471-2261-11-32
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Summary of included study characteristics
| Author (Year) | Type of scanner | Participants analysed (n) | Mean age (years) | Gender (M/F) | Prevalence of CAD (%) | Mean heart rate (± SD) (bpm) | Level of analysis presented in publication | ||
|---|---|---|---|---|---|---|---|---|---|
| Patient | Vessel | Segment | |||||||
| Brodoefel | Somatom Sensation 64, Siemens, Germany | 102 | 62.0 | 82/20 | 62.7 | 68.2 ± 13.3 | No | No | Yes |
| Cademartiri | Somatom Sensation 64, Siemens, Germany | 170 | 57.5 a | 124/46 | NR | 62.7 ± 10.5 | No | No | Yes |
| Ghostine | Somatom Sensation 64, Siemens, Germany | 93 | 65.0 | 61/32 | 46.0 | 73.0 ± 14.0 | Yes | No | Yes |
| Husmann | Somatom Sensation 64, Siemens, Germany | 88 | 64.3 | 48/40 | 48.9 | 63.0 ± 9.2 | Yes | Yes | Yes |
| Leber | Somatom Sensation 64, Siemens, Germany | 90 | 58.0 | 57/33 | 47.7 | 73.0 ± NR | Yes | No | Yes |
| Leschka | Somatom Sensation 64, Siemens, Germany | 74 | 61.7 | 50/24 | 47.3 | 67.7 ± 13.3 | Yes | Yes | Yes |
| Leschka | Somatom Sensation 64, Siemens, Germany | 114 | 62.2 | 73/41 | 62.3 | 68.0 ± 13.0 | Yes | No | Yes |
| Rixe | Somatom Sensation 64, Siemens, Germany | 76 | 65.5 | 47/29 | 52.6 | 68.0 ± 9.0 | Yes | Yes | Yes |
| Shabestari | Somatom Sensation 64, Siemens, Germany | 138 | 63.0 | 103/35 | 78.3 | 65.0 ± NR b | Yes | Yes | Yes |
| Achenbach | Somatom Sensation 64, Siemens Germany | 200 | 63.0 | 114/86 | 44.5 | 76 ± 13 | Yes | Yes | Yes |
| Budoff | Lightspeed VCT Scanner, GE Healthcare | 230 | 57.0 | 136/94 | 24.8 | 60 ± 12 | Yes | Yes | No |
| Cademartiri | Somatom Sensation 64, Siemens, Germany | 72 | 53.9 | 38/34 | 28.0 | 70.0 ± 9.9 | Yes | Yes | Yes |
| Hausleiter | Somatom Sensation 64 Cardiac, Siemens, Germany | 243 | 62.0 | 158 c/85 | 42.0 | 56.6 ± 6.5 | Yes | Yes | Yes |
| Herzog | Somatom Sensation 64, Siemens, Germany | 55 | 67.0 | 29/26 | 34.5 | 64.0 ± NR | Yes | Yes | Yes |
| Meijboom | Somatom Sensation 64, Siemens, Germany | 104 | 58.7 a | 75/29 | 85.0 | 66.0 ± 9.0 | Yes | Yes | Yes |
| Meijboom | Somatom Sensation 64; Siemens, Germany | 402 | 59.2 a | 279/123 | 62.9 | 59.5 a ± NR | Yes | Yes | Yes |
| Meijboom | Somatom Sensation 64, Siemens, Germany | 254 | 59.0 a | 171/83 | 49.6 | 59.3 a ± NR | Yes | Yes | Yes |
| Meijboom | Somatom Sensation 64, Siemens, Germany; Brilliance 64, Philips, The Netherlands; Toshiba Multi-Slice Aquilion 64, Toshiba, Japan | 360 | 60.0 | 245/115 | 68.0 | 59 ± 9 | Yes | Yes | Yes |
| Oncel | Somatom Sensation 64, Siemens, Germany | 80 | 56.0 | 61/19 | 77.5 | 58.0 ± 10.0 | Yes | No | Yes |
| Piers | Somatom Sensation 64, Siemens, Germany | 60 | 64.0 | 51/9 | 63.3 | 63.0 ± 12.0 | Yes | Yes | Yes |
| Scheffel | Somatom Sensation 64, Siemens, Germany | 120 | 68.2 | 71/49 | 55.0 | 59.0 ± 6.0 | Yes | Yes | Yes |
| Schlosser | Somatom Sensation 64, Siemens, Germany | 61 | 62.4 | 41/20 | NR | 57.0 ± 4.0 | No | Yes | Yes |
| Sheth | Toshiba Aquilion 64-detector scanner, Toshiba, Japan | 80 | 56.0 | 43/37 | 39.5 | NR | Yes | Yes | Yes |
| Weustink | DSCT Somatom Sensation 64, Siemens, Germany. | 100 | 61.0 | 79/21 | 77.0 | 68.0 ± 11.0 | Yes | Yes | Yes |
| Cademartiri | Somatom Sensation 64 Cardiac; Siemens, Germany | 134 | 63.4 | 98/36 | 62.7 | 57.5 a ± NR | Yes | Yes | Yes |
| Han | 64-slice VCT; GE Healthcare | 53 | 59.6 | 228 d/175 | 81.1 | NR | No | No | Yes |
| Pugliese | Somatom Sensation 64, Siemens, Germany | 51 | 59.0 | 39/12 | 74.5 | 58.0 ± 7.0 | Yes | No | Yes |
| Yoshida | Somatom Sensation 64, Siemens, Germany | 70 | 64.0 | 50/20 | NR | 65.0 ± 11.0 | No | No | Yes |
Bpm, beats-per-minute; CAD, coronary artery disease; F, female; M, male; NR, not reported.
Levels of evidence were defined according to the NHMRC diagnostic levels of evidence (11)
a Calculated post hoc using a crude weighted average of the two reported sub-groups
b Median heart rate
c There was a discrepancy between the number stated in Table 1 of the publication and the number stated in the text (228 and 226, respectively).
d For all enrolled patients in study
Figure 1Sensitivity and specificity for the base case meta-analysis at the patient level. CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive.
Figure 2PPV and NPV for the base case met-analysis at the patient level. CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive.
CTA diagnostic performance measures
| Analysis level | No. of included studies a | Sensitivity % (95% CI) | Specificity % (95% CI) | PPV Median (range) | NPV Median (range) | Diagnostic accuracy Median (range) |
|---|---|---|---|---|---|---|
| Patient: base case analysis | 18 | 98.2 (97.4-98.8) | 81.6 (79.0-84.0) | 90.5 (76-100) | 99.0 (83-100) | 92.0 (80-100) |
| Patient: alternative analysis | 22 | 98.0 (97.2-98.6) | 83.2 (81.1-85.2) | 89.0 (63-100) | 98.0 (83-100) | 92.0 (80-100) |
| Vessels: all | 17 | 94.9 (93.9-95.8) | 89.5 (88.8-90.2) | 75.0 (53-95) | 99.0 (93-100) | 91.5 (74-98) |
| RCA | 8 | 94.8 (92.0-96.9) | 91.0 (89.0-92.7) | 84.0 (73-94) | 98.5 (95-100) | 94.5 (84-99) |
| LM | 8 | 95.7 (85.2-99.5) | 97.1 (95.7-98.1) | 89.0 (24-100) | 100.0 (98-100) | 99.0 (91-100) |
| LAD | 7 | 97.4 (95.3-98.8) | 84.5 (82.1-86.7) | 78.0 (57-95) | 99.0 (95-100) | 93.0 (72-99) |
| CX | 8 | 94.1 (90.7-96.6) | 89.6 (87.7-91.3) | 78.5 (52-90) | 99.5 (95-100) | 94.0 (75-99) |
| Segments: all | 17 | 91.3 (90.2-92.2) | 94.0 (93.7-94.2) | 69.0 (44-86) | 99.0 (98-100) | 95.5 (90-99) |
CI, confidence interval; CX, circumflex; LAD, left anterior descending; LM, left main; NPV, negative predictive value; PPV, positive predictive value; RCA, right coronary artery
a Maximum number of included studies from which data were drawn to calculate diagnostic performance measures
Figure 3Summary receiver operator characteristic (SROC) curve for the base case patient level analysis.