Which pressure to believe? A comparison of direct arterial with indirect blood pressure measurement techniques in the pediatric intensive care unit.

Tanya R Holt1, Davinia E Withington, Ellen Mitchell.   

Abstract

OBJECTIVE: To determine the accuracy of arterial blood pressure monitoring using 1) direct arterial; 2) automated oscillometric; and 3) sphygmomanometer/Doppler ultrasound measurements in pediatric intensive care patients comparing methods 1) and 2) with 3), the gold standard used to define normal blood pressure.
DESIGN: Prospective observational study.
SETTING: Pediatric intensive care unit of a tertiary care pediatric teaching hospital. PATIENTS: Forty children (birth to 17 yrs) admitted to the pediatric intensive care unit with various clinical conditions requiring a radial arterial catheter for continuous arterial blood pressure monitoring.
INTERVENTIONS: None.
MEASUREMENTS AND MAIN RESULTS: Each subject had measurements taken every 6 hrs over a 24-hr period. Each set of measurements were: direct arterial blood pressure, indirect blood pressure using the Phillips automated oscillometric device, and indirect blood pressure using the sphygmomanometer and Doppler ultrasound. Analysis used the Bland-Altman plot followed by paired t testing to compare the three different methods. One hundred sixty triads of measurements were analyzed. There were no significant differences between the methods of blood pressure measurement when groups were analyzed based on age. When analyzed by age-specific normo-, hypo-, and hypertensive criteria, arterial blood pressure measurements agree closely with Doppler ultrasound readings, whereas systolic arterial blood pressure measurements were lower than indirect blood pressure using the Phillips automated oscillometric device readings in the hypotensive group (p < .001). In the hypertensive group, the systolic arterial blood pressure values were higher and indirect blood pressure using the Phillips automated oscillometric device readings lower (p < .001) than Doppler ultrasound (p = .03). There was no clinically significant difference between methods in the normotensive group. Diastolic blood pressure measurements were higher by arterial blood pressure in normotensive and hypertensive groups but no different in the hypotensive group.
CONCLUSION: Outside the normotensive range, the automated readings were higher during hypotension and lower during hypertension compared with the arterial and Doppler ultrasound methods. The arterial blood pressure was closer to the gold standard Doppler ultrasound blood pressure in all three blood pressure groups.

Entities:  

Mesh:

Year:  2011        PMID: 21666539     DOI: 10.1097/PCC.0b013e3182230f43

Source DB:  PubMed          Journal:  Pediatr Crit Care Med        ISSN: 1529-7535            Impact factor:   3.624


  13 in total

1.  Bupivacaine mandibular nerve block affects intraoperative blood pressure and heart rate in a Yucatan miniature swine mandibular condylectomy model: a pilot study.

Authors:  Jonathan F Bova; Anderson F da Cunha; Rhett W Stout; Sarindr Bhumiratana; David M Alfi; Sidney B Eisig; Gordana Vunjak-Novakovic; Mandi J Lopez
Journal:  J Invest Surg       Date:  2014-11-13       Impact factor: 2.533

2.  Hypertension and health outcomes in the PICU.

Authors:  Brett J Ehrmann; David T Selewski; Jonathan P Troost; Susan M Hieber; Debbie S Gipson
Journal:  Pediatr Crit Care Med       Date:  2014-06       Impact factor: 3.624

Review 3.  Evaluation and management of elevated blood pressures in hospitalized children.

Authors:  Abanti Chaudhuri; Scott M Sutherland
Journal:  Pediatr Nephrol       Date:  2018-08-31       Impact factor: 3.714

4.  Blood pressure monitoring following kidney transplantation in children: a comparison of invasive and noninvasive measurements using Doppler as a benchmark technique.

Authors:  Eytan Kaplan; Gili Kadmon; Elhanan Nahum; Hadas Alfandary; Orly Haskin; Avichai Weissbach
Journal:  Pediatr Nephrol       Date:  2022-08-01       Impact factor: 3.651

5.  Risk of over-diagnosis of hypotension in children: a comparative analysis of over 50,000 blood pressure measurements.

Authors:  Samiran Ray; Libby Rogers; David P Noren; Ranjana Dhar; Simon Nadel; Mark J Peters; David P Inwald
Journal:  Intensive Care Med       Date:  2017-05-19       Impact factor: 17.440

6.  Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU.

Authors:  Sara Ribezzo; Eleonora Spina; Stefano Di Bartolomeo; Gianfranco Sanson
Journal:  ScientificWorldJournal       Date:  2014-01-30

7.  Accuracy in Blood Pressure Monitoring: The Effect of Noninvasive Blood Pressure Cuff Inflation on Intra-arterial Blood Pressure Values.

Authors:  Veena Sheshadri; Akhilesh Kumar Tiwari; Mahesh Nagappa; Lashmi Venkatraghavan
Journal:  Anesth Essays Res       Date:  2017 Jan-Mar

Review 8.  Advances in Non-Invasive Blood Pressure Monitoring.

Authors:  Xina Quan; Junjun Liu; Thomas Roxlo; Siddharth Siddharth; Weyland Leong; Arthur Muir; So-Min Cheong; Anoop Rao
Journal:  Sensors (Basel)       Date:  2021-06-22       Impact factor: 3.576

9.  Identification of hypertension in hospitalized children prescribed as-needed antihypertensive medication.

Authors:  Kathryn R Kocher; Dmitry Tumin; Amber G Lehmann; Lilliana Michelle Gomez Mendez
Journal:  J Clin Hypertens (Greenwich)       Date:  2020-08-04       Impact factor: 3.738

10.  The accuracy of blood pressure measured by arterial line and non-invasive cuff in critically ill children.

Authors:  Rachel Joffe; Jonathan Duff; Gonzalo Garcia Guerra; Jodie Pugh; Ari R Joffe
Journal:  Crit Care       Date:  2016-06-08       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.