Literature DB >> 21664841

The effect of aging on luminance of standard liquid crystal display (LCD) monitors.

Kristina Hellén-Halme1, Bengt Hellén-Halme, Ann Wenzel.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the luminance level of monitors as a function of the burning time of the lamp in the monitor and further to evaluate if different areas of the monitor varied in luminance levels at a specific time.
MATERIAL AND METHODS: Nine standard liquid crystal display (LCD) monitor models, in total 180 monitors, were evaluated. The monitors' maximal luminance levels were measured, and the burning times of the lamps were noted. The estimated function between burning time and luminance was calculated. On 11 monitors, 9 areas distributed over the full screen were measured to evaluate the extent to which different areas had different luminance levels.
RESULTS: Over time, the luminance decreased in all monitor models. No difference was found among the monitor models in how rapidly the luminance levels decreased. Different luminance values were found for different areas on the monitors, but no area deteriorated faster than another.
CONCLUSIONS: Standard liquid crystal display monitors undergo gradual deteriorations in luminance levels over time, and these deteriorations are correlated with the time that the monitors have been switched on. The degradation in luminance is similar over all areas of the monitors.
Copyright © 2011 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21664841     DOI: 10.1016/j.tripleo.2011.02.044

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  5 in total

1.  Effective luminance deterioration of medical liquid crystal displays in clinical use.

Authors:  Keita Takahashi; Shinichi Awamoto; Shinya Takarabe; Kazuhisa Ogawa; Yasuhiko Nakamura
Journal:  Radiol Phys Technol       Date:  2017-05-02

2.  Viewing your digital radiographs: which monitor is best?

Authors:  D McIlgorm
Journal:  Br Dent J       Date:  2016-04-22       Impact factor: 1.626

3.  The effect of time in use on the display performance of the iPad.

Authors:  Liam J Caffery; Kenneth L Manthey; Lawrence H Sim
Journal:  Br J Radiol       Date:  2016-05-16       Impact factor: 3.039

4.  A method for evaluating luminance non-uniformity of displays by use of a commercially available digital camera.

Authors:  Keishin Kawamoto; Shogo Tokurei; Junji Morishita
Journal:  Radiol Phys Technol       Date:  2017-08-16

5.  Impact of viewing conditions on the performance assessment of different computer monitors used for dental diagnostics.

Authors:  Thomas Hastie; Sascha Venske-Parker; Johan K M Aps
Journal:  Imaging Sci Dent       Date:  2021-02-09
  5 in total