A M Kamel1, P R Brown, B Munson.
Abstract
The effects of mobile-phase additives and analyte concentration on electrospray ionization mass spectra of a series of tetracyclines were investigated in both positive and negative ion modes. Only [M + H](+) and [M - H](-) ions were observed. The greatest sensitivity as [M + H](+) ions was obtained with 1% acetic acid and the greatest sensitivity as [M - H](-) ions was obtained using 50 mM ammonium hydroxide. Sensitivities in the positive ion mode were greater than those in the negative ion mode. The sensitivity as [M + H](+) showed no systematic variation with pH; however, the sensitivity as [M - H](-) did increase with increasing pH. A larger linear range was observed for [M - H](-) than for [M + H](+) ions. Both [M + Na](+) and [M + H](+) ions were observed with 0.5 mM sodium acetate and sodium iodide, but no adduct ions were observed with ammonium acetate. Some M(2)H(+) ions were observed at higher concentrations. Cluster ions, Na(NaOAc)(n)(+) or Na(NaI)(n)(+), but no sample ions were observed using 5 mM salts. The data suggest that mechanisms in addition to solution ionization are involved in the formation of the ESI sample ions. The utility of mobile phases containing 1% HOAc or 50 mM NH(4)OH was demonstrated for chromatographic separations.Entities:
Year: 1999 PMID: 21662767 DOI: 10.1021/ac9807114