Literature DB >> 21656456

A new large-caliber trocar for percutaneous endoscopic gastrostomy by the introducer technique in head and neck cancer patients.

J H Giordano-Nappi1, F Maluf-Filho, S Ishioka, F Y Hondo, S E Matuguma, M Simas de Lima, M Lera dos Santos, F A Retes, P Sakai.   

Abstract

BACKGROUND AND STUDY AIMS: In many patients, percutaneous endoscopic gastrostomy (PEG) can be limited by digestive tract stenosis. PEG placement using an introducer is the safest alternative for this group of patients, but the available devices are difficult to implement and require smaller-caliber tubes. The aim of this study was to evaluate the modification of an introducer technique device for PEG placement with regard to the following: procedure feasibility, possibility of using a 20-Fr balloon gastrostomy tube, tube-related function and problems, complications, procedure safety, and mortality. PATIENTS AND METHODS: Between March 2007 and February 2008, 30 consecutive patients with head and neck malignancies underwent introducer PEG placement with the modified device and gastropexy. Each patient was evaluated for 60 days after the procedure for the success of the procedure, infection, pain, complications, mortality, and problems with the procedure.
RESULTS: The procedure was successful in all cases with no perioperative complications. No signs of stomal infection were observed using the combined infection score. The majority of patients experienced mild-to-moderate pain both in the immediate postoperative period and at 72 hours. One major early complication (3.3%) and two minor complications (6.7%) were observed. No procedure-related deaths occurred during the first 60 days after the procedure.
CONCLUSION: The device modification for PEG using the introducer technique is feasible, safe, and efficient in outpatients with obstructive head and neck cancer. In this series, it allowed the use of a larger-caliber tube with low complication rates and no procedure-related mortality. © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Year:  2011        PMID: 21656456     DOI: 10.1055/s-0030-1256495

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  6 in total

1.  A novel percutaneous magnetically guided gastrostomy technique without endoscopy or imaging guidance: a feasibility study in a porcine model.

Authors:  Eduardo Aimore Bonin; Paulo Roberto Walter Ferreira; Marcelo de Paula Loureiro; Thais Andrade Costa-Casagrande; Paolo de Oliveira Salvalaggio; Guilherme Francisco Gomes; Rafael William Noda; Christopher John Gostout; Leandro Totti Cavazzola
Journal:  Surg Endosc       Date:  2018-07-18       Impact factor: 4.584

2.  Comparison of the pull and introducer percutaneous endoscopic gastrostomy techniques in patients with head and neck cancer.

Authors:  Felipe A Retes; Fabio S Kawaguti; Marcelo S de Lima; Bruno da Costa Martins; Ricardo S Uemura; Gustavo A de Paulo; Caterina Mp Pennacchi; Carla Gusmon; Adriana Vs Ribeiro; Elisa R Baba; Sebastian N Geiger; Mauricio P Sorbello; Marco A Kulcsar; Ulysses Ribeiro; Fauze Maluf-Filho
Journal:  United European Gastroenterol J       Date:  2016-07-21       Impact factor: 4.623

3.  Complications of percutaneous endoscopic gastrostomy tube insertion in cancer patients: a retrospective study.

Authors:  Hala Mansoor; Muhammad Adnan Masood; Muhammed Aasim Yusuf
Journal:  J Gastrointest Cancer       Date:  2014-12

Review 4.  Usefulness of percutaneous endoscopic gastrostomy for supportive therapy of advanced aerodigestive cancer.

Authors:  Haruei Ogino; Hirotada Akiho
Journal:  World J Gastrointest Pathophysiol       Date:  2013-11-15

5.  Comparison of Pull and Introducer Techniques for Percutaneous Endoscopic Gastrostomy.

Authors:  Xia Li; Jun-Xi Wang; Yu-Ping Wang; Jia-Xin Shen; Yi-Xing Zheng; Pei-Hong Zhang; Jing-Jing Wei; Ze-Hao Zhuang
Journal:  J Multidiscip Healthc       Date:  2022-04-05

6.  Percutaneous endoscopic gastrostomy (PEG) using a novel large-caliber introducer technique kit: a retrospective analysis.

Authors:  Ezekiel Wong Toh Yoon; Kaori Yoneda; Shinya Nakamura; Kazuki Nishihara
Journal:  Endosc Int Open       Date:  2016-08-30
  6 in total