Literature DB >> 21637120

Biomechanical evaluation of two-part surgical neck fractures of the humerus fixed by an angular stable locked intramedullary nail.

Joachim Horn1, Boyko Gueorguiev, Stefano Brianza, Harald Steen, Karsten Schwieger.   

Abstract

OBJECTIVE: The aim of the current study was to see how different interlocking mechanisms would affect construct stability and overall failure in the treatment of two-part surgical neck fractures in the proximal humerus in vitro.
METHODS: Left and right bones of eight pairs of fresh-frozen human cadaveric humeri were assigned to either a group with conventional or a group with angular stable distal interlocking. The different experimental interlocking mechanisms were used in a surgical neck fracture model of the humerus (Orthopaedic Trauma Association 11- A3) stabilized by a proximal humeral nail. The following variables were evaluated by biomechanical tests: hysteresis width in bending and torsion, stiffness, and fracture gap movement during cyclic axial loading until failure and the overall failure mechanism of the construct.
RESULTS: The angular stable group showed significantly less motion in initial bending and torsion and higher bending stiffness throughout the complete deformation cycle compared with the conventional interlocked group. Fracture gap movement was significantly less in the angular stable group. Higher stability was mainly observed in the early phase of the applied loading pattern; however, ultimate failure was not related to distal interlocking but occurred in the proximal fragment in both groups.
CONCLUSIONS: An experimental angular stable distal interlocking system of proximal humeral nails shows higher construct stability in the early phase of fracture fixation in vitro. In terms of overall failure, loss of fixation in the proximal fragment was crucial and not different between groups.

Entities:  

Mesh:

Year:  2011        PMID: 21637120     DOI: 10.1097/BOT.0b013e3181ffdf9a

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  Internal fixation of fractures of the proximal humerus with the MultiLoc nail.

Authors:  M H Hessmann; S Nijs; T Mittlmeier; M Kloub; M J M Segers; V Winkelbach; M Blauth
Journal:  Oper Orthop Traumatol       Date:  2012-09       Impact factor: 1.154

2.  Operative treatment of 2-part surgical neck fractures of the proximal humerus (AO 11-A3) in the elderly: Cement augmented locking plate Philos™ vs. proximal humerus nail MultiLoc®.

Authors:  Tobias Helfen; Georg Siebenbürger; Marcel Mayer; Wolfgang Böcker; Ben Ockert; Florian Haasters
Journal:  BMC Musculoskelet Disord       Date:  2016-10-28       Impact factor: 2.362

3.  Is Bridge Plating of Comminuted Humeral Shaft Fractures Advantageous When Using Compression Plates with Three versus Two Screws per Fragment? A Biomechanical Cadaveric Study.

Authors:  Guilherme Seva Gomes; Ivan Zderic; Marc-Daniel Ahrend; Kodi E Kojima; Peter Varga; William Dias Belangero; Geoff Richards; Simon M Lambert; Boyko Gueorguiev
Journal:  Biomed Res Int       Date:  2021-03-06       Impact factor: 3.411

  3 in total