Transcutaneous regional venous oximetry: a feasibility study.

Robert H Thiele1, Jason M Tucker-Schwartz, Yao Lu, George T Gillies, Marcel E Durieux.   

Abstract

BACKGROUND: The arterial pulse oximeter, which was introduced clinically in the 1970s, is a convenient, useful, and now ubiquitous anesthesia monitor. Unfortunately, although percent saturation of arterial hemoglobin is, along with cardiac output and concentration of hemoglobin, one of 3 components of oxygen delivery, it does not indicate whether oxygen delivery to a region of interest is adequate. Knowledge of peripheral or regional venous oxygen saturation (Sxvo₂) may lend insight into analysis of regional oxygen supply and demand. Our goal was to assess the suitability of 3 anatomic sites for the transcutaneous assessment of Sxvo₂.
METHODS: Using a Nonin reflectance oximetry probe (provided by Nonin Medical, Plymouth, MN) placed directly over the antecubital, external jugular, and internal jugular veins in 10 volunteers, we measured the absorbance of red and infrared electromagnetic radiation. We performed fast Fourier transformation on these absorbance waveforms. The ratio of pulsatile absorbance of red and infrared radiation at different frequencies was compared with nonpulsatile absorption, and Sxvo₂ was calculated based on previously derived empiric correlations.
RESULTS: Estimates of transcutaneous Sxvo₂ ranged from 41% to 97%, with mean values of 75%, 80%, and 80% at the antecubital, external jugular, and internal jugular veins, respectively. Overall, 93% of predicted Sxvo₂ values were < 90%.
CONCLUSION: Validation and subsequent improvement of this technique requires correlation of our results with venous blood gas measurements, followed by incorporation of technologies from related fields in oximetry (fetal reflectance oximetry and near-infrared spectroscopy), as well as the development of advanced signal processing techniques.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21613200     DOI: 10.1213/ANE.0b013e3182166a26

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  6 in total

1.  Frequency domain analysis of cerebral near infrared spectroscopy signals during application of an impedance threshold device in spontaneously ventilating volunteers.

Authors:  Douglas A Colquhoun; Kimberly Naden; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2015-06-27       Impact factor: 2.502

2.  Measuring blood oxygen saturation along a capillary vessel in human.

Authors:  Kfir Akons; Eldad J Dann; Dvir Yelin
Journal:  Biomed Opt Express       Date:  2017-10-30       Impact factor: 3.732

3.  Non-invasive estimation of jugular venous oxygen saturation: a comparison between near infrared spectroscopy and transcutaneous venous oximetry.

Authors:  Douglas A Colquhoun; Jason M Tucker-Schwartz; Marcel E Durieux; Robert H Thiele
Journal:  J Clin Monit Comput       Date:  2012-04       Impact factor: 2.502

Review 4.  Respiratory physiology and the impact of different modes of ventilation on the photoplethysmographic waveform.

Authors:  Aymen A Alian; Kirk H Shelley
Journal:  Sensors (Basel)       Date:  2012-02-16       Impact factor: 3.576

5.  Comparison of non-invasive peripheral venous saturations with venous blood co-oximetry.

Authors:  A M Belhaj; J P Phillips; P A Kyriacou; R M Langford
Journal:  J Clin Monit Comput       Date:  2016-11-21       Impact factor: 2.502

Review 6.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.