Literature DB >> 21606264

A phantom study comparing ultrasound-guided liver tumor puncture using new real-time 3D ultrasound and conventional 2D ultrasound.

Katsutoshi Sugimoto1, Fuminori Moriyasu, Junji Shiraishi, Masahiko Yamada, Yasuharu Imai.   

Abstract

OBJECTIVE: The purpose of this study is to compare the accuracy of ultrasound-guided puncture using new real-time 3D (4D) ultrasound and conventional 2D ultrasound for focal hepatic masses using a liver phantom.
MATERIALS AND METHODS: A 4D ultrasound system equipped with a 5-MHz 4D probe displayed both axial and orthogonal images parallel to a puncture line plane. We used a liver phantom that contained four simulated spherical masses in an acrylic box (length × width × height, 300 × 299 × 150 mm) with two different sizes (15 and 30 mm in diameter) in two different positions (30 and 80 mm from the surface). Four inexperienced and four experienced physicians attempted punctures on these four simulated masses twice using 2D and 4D ultrasound guidance in a total of 128 punctures (eight operators, two techniques, and eight punctures per session). The error distance of the puncture was defined as the perpendicular distance from the center of a target mass (sphere) to the line of the puncture needle in the coronal plane of the target center, which was measured manually on the basis of the 3D volume data on off-line analysis.
RESULTS: On each tumor model, the average error distance with 4D ultrasound was significantly smaller than that with 2D ultrasound, except for one tumor model that was 15 mm in diameter and 30 mm in depth. The average error distances for the experienced group tended to be smaller than those for the inexperienced group, with both techniques and on each tumor model, and there was a statistically significant difference between the two groups on one tumor model (30 mm in diameter and 80 mm in depth) on 4D ultrasound (p < 0.05).
CONCLUSION: Four-dimensional ultrasound-guided puncture for liver tumors can markedly improve puncture accuracy for both experienced and inexperienced physicians compared with conventional 2D ultrasound guidance.

Entities:  

Mesh:

Year:  2011        PMID: 21606264     DOI: 10.2214/AJR.10.5552

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  7 in total

1.  Application of real-time three-dimensional contrast-enhanced ultrasound using SonoVue for the evaluation of focal liver lesions: a prospective single-center study.

Authors:  Ying Lu; Baoxian Liu; Yanling Zheng; Jia Luo; Xiaoer Zhang; Guangliang Huang; Xiaohua Xie; Jieyi Ye; Wei Wang; Xusheng Liu; Xiaoyan Xie
Journal:  Am J Transl Res       Date:  2018-05-15       Impact factor: 4.060

Review 2.  Development of technologies for placement of perineural catheters.

Authors:  Hesham Elsharkawy; Ankit Maheshwari; Ehab Farag; Edward R Mariano; Richard W Rosenquist
Journal:  J Anesth       Date:  2015-09-14       Impact factor: 2.078

3.  Development and evaluation of an original phantom model of ultrasonography-guided thyroid gland biopsy for the training of surgical residents and students.

Authors:  Masayuki Baba; Keitaro Matsumoto; Hisakazu Shindo; Megumi Matsumoto; Ryota Otsubo; Aya Tanaka; Shosaburo Oyama; Rui Zhu; Ikuo Yamamoto; Takeshi Nagayasu
Journal:  Surg Today       Date:  2022-10-01       Impact factor: 2.540

4.  Volume navigation with contrast enhanced ultrasound and image fusion for percutaneous interventions: first results.

Authors:  Ernst Michael Jung; Chris Friedrich; Patrick Hoffstetter; Lena Marie Dendl; Frank Klebl; Ayman Agha; Phillipp Wiggermann; Christian Stroszcynski; Andreas Georg Schreyer
Journal:  PLoS One       Date:  2012-03-20       Impact factor: 3.240

5.  Anthropomorphic liver phantom with flow for multimodal image-guided liver therapy research and training.

Authors:  Anna Rethy; Jørn Ove Sæternes; Jostein Halgunset; Ronald Mårvik; Erlend F Hofstad; Juan A Sánchez-Margallo; Thomas Langø
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-09-19       Impact factor: 2.924

6.  Dynamic Hepatocellular Carcinoma Model Within a Liver Phantom for Multimodality Imaging.

Authors:  Muntaser S Ahmad; Nursakinah Suardi; Ahmad Shukri; Nik Noor Ashikin Nik Ab Razak; Ammar A Oglat; Osama Makhamrah; Hjouj Mohammad
Journal:  Eur J Radiol Open       Date:  2020-09-03

7.  A standardized method for 4D ultrasound-guided peripheral nerve blockade and catheter placement.

Authors:  N J Clendenen; C B Robards; S R Clendenen
Journal:  Biomed Res Int       Date:  2014-01-19       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.