Assessment of ultrasound monitor image display performance.

Sally C Moore1, Craig R Munnings, David S Brettle, J Anthony Evans.   

Abstract

The display monitor on an ultrasound scanner is used to make primary diagnoses. In this study, 31 ultrasound systems were assessed against current American Association of Physicists in Medicine (AAPM) display standards. Measurements of peak levels (L(max) and L(min)) were generated. Ambient light, L(amb) (cd/m(2)) and room illuminance, L(x) (Lux) were measured. Luminance ratio was calculated (LR' = (L(max)+L(amb))/(L(min)+L(amb))). Initially, only 8/31 systems (26%) passed all the criteria. After adjustment, a further 7/31 (23%) passed making a total of 15/31 passes (48%). A total of 16/31 (52%) were considered overall fails: three due to poor room lighting, 14 due to poor monitor performance. Considering errors this could be as low as 6/31 (19%). Although further work is required to confirm the applicability of these results, it is of concern that three-quarters of ultrasound scanners could be suboptimally adjusted with 19%-55% unable to pass the AAPM criteria. The impact of this on clinical practice is unknown but there is clearly a need to review display quality assurance on ultrasound scanners.
Copyright © 2011 World Federation for Ultrasound in Medicine & Biology. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2011        PMID: 21601138     DOI: 10.1016/j.ultrasmedbio.2011.02.018

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  2 in total

1.  Reliability of aortic aneurysm screening measurements.

Authors:  Rachel Weston Smith; Tony Evans; Stephen Wolstenhulme
Journal:  Ultrasound       Date:  2013-12-13

2.  Agreement between objective and subjective assessment of image quality in ultrasound abdominal aortic aneurism screening.

Authors:  S Wolstenhulme; A G Davies; C Keeble; S Moore; J A Evans
Journal:  Br J Radiol       Date:  2014-12-11       Impact factor: 3.039

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.