Repeatability of the ciliary sulcus-to-sulcus diameter measurement using wide-scanning-field ultrasound biomicroscopy.

Sho Yokoyama1, Takashi Kojima, Rie Horai, Mayuka Ito, Tomoaki Nakamura, Kazuo Ichikawa.   

Abstract

PURPOSE: To evaluate the interexaminer and intraexaminer variance in ciliary sulcus-to-sulcus (STS) diameter measurements using wide-scanning-field ultrasound biomicroscopy (UBM) (Vumax II).
SETTING: Nagoya Eye Clinic, Nagoya, Japan.
DESIGN: Evaluation of diagnostic test or technology.
METHOD: Two examiners measured the STS diameter 10 times in each subject. Intraexaminer and interexaminer variances were calculated. To determine interexaminer variance at each UBM measurement step, the examiner was changed every 2 steps when the STS diameter was measured. The Bland-Altman plot test was used to analyze each step of measurement variance.
RESULTS: Thirty eyes of 15 volunteers (9 men, 6 women; mean age 29.6 years ± 6.2 [SD]) were enrolled. The mean coefficient of variation was 0.62% ± 0.20% for intraexaminer measurements and 3.4% ± 2.4% for interexaminer measurements. There were statistically significant differences between the 2 examiners in the step of extracting images and determining and measuring the STS diameter (P=.02). The 95% confidence intervals in all steps were relatively high (recordings, -1.02 to 1.08 mm; extracting images and determining and measuring STS; -0.79 to 1.25 mm).
CONCLUSION: Interexaminer variance in STS measurement should be considered if the STS diameter is used for selecting the size of posterior chamber phakic intraocular lenses.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21570248     DOI: 10.1016/j.jcrs.2011.01.024

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

Review 1.  Test-retest variability in structural parameters measured with glaucoma imaging devices.

Authors:  Makoto Araie
Journal:  Jpn J Ophthalmol       Date:  2012-11-09       Impact factor: 2.447

Review 2.  Meta-analysis and review: effectiveness, safety, and central port design of the intraocular collamer lens.

Authors:  Mark Packer
Journal:  Clin Ophthalmol       Date:  2016-06-09

3.  Adjustment formulae to improve the correlation of white-to-white measurement with direct measurement of the ciliary sulcus diameter by ultrasound biomicroscopy.

Authors:  Seyed Javad Hashemian; Majid Mohebbi; Mehdi Yaseri; Mohammad Ebrahim Jafari; Shadrokh Nabili; Seyed Mahyar Hashemian; Mahsa Sadat Hashemian
Journal:  J Curr Ophthalmol       Date:  2017-12-16

4.  Phakic intraocular lens: Getting the right size.

Authors:  Kalyaani Deshpande; Rushad Shroff; Partha Biswas; Kamal Kapur; Naren Shetty; Ann Sarah Koshy; Pooja Khamar
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

5.  Agreement between the biometric measurements used to calculate the size of the implantable collamer lenses measured with four different technologies.

Authors:  Jorge A Calvo-Sanz; Francisco Poyales; Ying Zhou; Alfonso Arias-Puente; Nuria Garzón
Journal:  Indian J Ophthalmol       Date:  2022-05       Impact factor: 2.969

Review 6.  Anterior segment imaging in glaucoma: An updated review.

Authors:  Jessica S Maslin; Yaniv Barkana; Syril K Dorairaj
Journal:  Indian J Ophthalmol       Date:  2015-08       Impact factor: 1.848

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.