Abstract
An electrochemical immunosensing assay for myeloperoxidase (MPO) determination in human serum has been developed. Firstly, L-Cysteine was initially electropolymerized on an Au electrode to form L-Cysteine film. After that cerium dioxide (CeO2) dispersed in 1-butyl-3-methylimidazolium hexafluorophosphate (BMIMPF6) were immobilized on the L-Cysteine film. Then the negatively charged nanogold particles were adsorbed onto the membrane via the positive charge of CeO2, which aimed at assembling more antibody of MPO (anti-MPO). The resulting immunosensor showed a high sensitivity, broad linear response to the MPO concentration comprised between 10 ng/mL and 400 ng/mL with a detection limit of 0.06 ng/mL. Moreover, the surface morphology of the electrode was studied by means of a scanning electron microscope and the electrochemical properties of the fabricated immunosensor were further characterized by cyclic voltammetry. Also, factors influencing the performance of the resulting immunosensors were studied in detail.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21561744 DOI: 10.1016/j.colsurfb.2011.04.017
Source DB: PubMed Journal: Colloids Surf B Biointerfaces ISSN: 0927-7765 Impact factor: 5.268