[Clinical experience with the Kensey catheter system--complications and results].

V H Schild1, G Zocholl, H Kopp, W Schmiedt, K Schunk, U Hake, H Jakob, M Thelen.   

Abstract

Fifteen procedures have been carried out in 14 patients using the Kensey rotation angioplasty system. Sixteen occlusions of the superficial femoral artery, with an average length of 10.2 cm (3 to 24 cm) and one artery with multiple stenoses were treated. The following results were obtained: creation of a lumen, with subsequent balloon dilatation in eight out of fifteen procedures; in two patients the vessels were occluded shortly after the procedure, due to complications (one obstructive plaque and one haematoma); five perforations, three dissections and one peripheral embolus. The total complication rate was 11 out of 15 cases, clinical improvement was obtained in six out of 15 cases.

Entities:  

Mesh:

Year:  1990        PMID: 2155449     DOI: 10.1055/s-2008-1046845

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  2 in total

1.  Percutaneous thrombectomy of the acutely thrombosed dialysis graft: in vitro evaluation of four devices.

Authors:  T Schmitz-Rode; J G Pfeffer; K Bohndorf; R W Günther
Journal:  Cardiovasc Intervent Radiol       Date:  1993 Mar-Apr       Impact factor: 2.740

2.  Distribution and size of particles after dynamic angioplasty in cadaveric arteries with the Kensey catheter system: a comparison with clinical experience.

Authors:  D Moellmann; F P Kuhn; D Bömer
Journal:  Cardiovasc Intervent Radiol       Date:  1992 Jul-Aug       Impact factor: 2.740

  2 in total