Evaluation of a rotary laser body scanner for body volume and fat assessment.

M Reese Pepper1, Jeanne H Freeland-Graves, Wurong Yu, Philip R Stanforth, Bugao Xu.   

Abstract

This paper reports the evaluation tests on the reliability and validity of a 3-dimensional (3D) laser body scanner for estimation of body volume and % fat. Repeated measures of body imaging were performed for reproducibility analysis. Validity of the instrument was assessed by comparison of measures of body volume by imaging to hydrodensitometry, and body fat was compared to hydrodensitometry and dual energy X-ray absorptiometry. Reproducibility analysis showed little difference between within-subjects measurements of volume (ICC ≥ 0.99, p < 0.01). Body volume estimations by laser body scanner and hydrodensitometry were strongly related (r = 0.99, p < 0.01), and agreement was high (ICC = 0.99, p < 0.01). Measurements of % body fat also agreed strongly with each other between methods (ICC = 0.86, p < 0.01), and mean % fat estimates by body imaging did not differ from criterion methods (p > 0.05). These findings indicate that the 3D laser body scanner is a reliable and valid technique for the estimation of body volume. Furthermore, body imaging is an accurate measure of body fat, as compared to dual energy X-ray absorptiometry. This new instrument is promising as a quick, simple to use, and inexpensive method of body composition analysis.

Entities:  

Year:  2010        PMID: 21552454      PMCID: PMC3087194          DOI: 10.1520/JTE102871

Source DB:  PubMed          Journal:  J Test Eval        ISSN: 0090-3973            Impact factor:   1.264


  23 in total

1.  Implementation of whole body scanner for determining somatotype index at Chang Gung Memorial Hospital.

Authors:  T H Liu; W K Chiou; J D Lin; C Y Yu
Journal:  Chang Gung Med J       Date:  2001-11

2.  The gross composition of the body.

Authors:  W E SIRI
Journal:  Adv Biol Med Phys       Date:  1956

3.  Predictive validity of four bioelectrical impedance equations in determining percent fat mass in overweight and obese children.

Authors:  Jane Cleary; Suzie Daniells; Anthony D Okely; Marijka Batterham; Jessie Nicholls
Journal:  J Am Diet Assoc       Date:  2008-01

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Comparison of three-dimensional anthropometric body surface scanning to waist-hip ratio and body mass index in correlation with metabolic risk factors.

Authors:  Jen-Der Lin; Wen-Ko Chiou; Hsiao-Fen Weng; Ying-Huang Tsai; Thu-Hua Liu
Journal:  J Clin Epidemiol       Date:  2002-08       Impact factor: 6.437

6.  Regional body volumes, BMI, waist circumference, and percentage fat in severely obese adults.

Authors:  Jack Wang; Dympna Gallagher; John C Thornton; Wen Yu; Rich Weil; Betty Kovac; F Xavier Pi-Sunyer
Journal:  Obesity (Silver Spring)       Date:  2007-11       Impact factor: 5.002

7.  Age-variability in body shape associated with excess weight: the UK National Sizing Survey.

Authors:  Jonathan C K Wells; Tim J Cole; Philip Treleaven
Journal:  Obesity (Silver Spring)       Date:  2008-02       Impact factor: 5.002

8.  Lung volumes in healthy nonsmoking adults.

Authors:  R O Crapo; A H Morris; P D Clayton; C R Nixon
Journal:  Bull Eur Physiopathol Respir       Date:  1982 May-Jun

9.  BMI compared with 3-dimensional body shape: the UK National Sizing Survey.

Authors:  Jonathan C K Wells; Philip Treleaven; Tim J Cole
Journal:  Am J Clin Nutr       Date:  2007-02       Impact factor: 7.045

10.  Measuring body composition in overweight individuals by dual energy x-ray absorptiometry.

Authors:  Rhonda A Brownbill; Jasminka Z Ilich
Journal:  BMC Med Imaging       Date:  2005-03-04       Impact factor: 1.930

View more
  3 in total

Review 1.  Multi-component molecular-level body composition reference methods: evolving concepts and future directions.

Authors:  S B Heymsfield; C B Ebbeling; J Zheng; A Pietrobelli; B J Strauss; A M Silva; D S Ludwig
Journal:  Obes Rev       Date:  2015-02-03       Impact factor: 9.213

2.  Prediction of Android and Gynoid Body Adiposity via a Three-dimensional Stereovision Body Imaging System and Dual-Energy X-ray Absorptiometry.

Authors:  Jane J Lee; Jeanne H Freeland-Graves; M Reese Pepper; Philip R Stanforth; Bugao Xu
Journal:  J Am Coll Nutr       Date:  2015-04-27       Impact factor: 3.169

3.  Validity and reliability of total body volume and relative body fat mass from a 3-dimensional photonic body surface scanner.

Authors:  Carolin Adler; Astrid Steinbrecher; Lina Jaeschke; Anja Mähler; Michael Boschmann; Stephanie Jeran; Tobias Pischon
Journal:  PLoS One       Date:  2017-07-03       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.