AIM: To determine interfractional three-dimensional set-up errors using X-ray volumetric imaging (XVI).
MATERIALS AND METHODS: Between December 2007 and August 2009, 125 patients were taken up for image-guided radiotherapy using online XVI. After matching of reference and acquired volume view images, set-up errors in three translation directions were recorded and corrected online before treatment each day. Mean displacements, population systematic (Σ), and random (σ) errors were calculated and analyzed using SPSS (v16) software. Optimum clinical target volume (CTV) to planning target volume (PTV) margin was calculated using Van Herk's (2.5Σ + 0.7 σ) and Stroom's (2Σ + 0.7 σ) formula.
RESULTS: Patients were grouped in 4 cohorts, namely brain, head and neck, thorax, and abdomen-pelvis. The mean vector displacement recorded were 0.18 cm, 0.15 cm, 0.36 cm, and 0.35 cm for brain, head and neck, thorax, and abdomen-pelvis, respectively. Analysis of individual mean set-up errors revealed good agreement with the proposed 0.3 cm isotropic margins for brain and 0.5 cm isotropic margins for head-neck. Similarly, 0.5 cm circumferential and 1 cm craniocaudal proposed margins were in agreement with thorax and abdomen-pelvic cases.
CONCLUSION: The calculated mean displacements were well within CTV-PTV margin estimates of Van Herk (90% population coverage to minimum 95% prescribed dose) and Stroom (99% target volume coverage by 95% prescribed dose). Employing these individualized margins in a particular cohort ensure comparable target coverage as described in literature, which is further improved if XVI-aided set-up error detection and correction is used before treatment.

Entities:  

Mesh:

Year:  2011        PMID: 21546741     DOI: 10.4103/0973-1482.80457

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  4 in total

1.  Comparison of set-up errors by breast size on wing board by portal imaging.

Authors:  Waseem Raza; Sushma Agarwal; Koilpillai Joseph Maria Das; Sellepolyam Kaliyaperumal Senthil Kumar; Punita Lal
Journal:  Rep Pract Oncol Radiother       Date:  2016-05-06

Review 2.  Image guidance in radiation therapy: techniques and applications.

Authors:  Shikha Goyal; Tejinder Kataria
Journal:  Radiol Res Pract       Date:  2014-12-17

3.  Analysis of the Setup Uncertainty and Margin of the Daily ExacTrac 6D Image Guide System for Patients with Brain Tumors.

Authors:  Se An Oh; Ji Woon Yea; Min Kyu Kang; Jae Won Park; Sung Kyu Kim
Journal:  PLoS One       Date:  2016-03-28       Impact factor: 3.240

4.  Analysis of setup error based on CTVision for nasopharyngeal carcinoma during IGRT.

Authors:  Shuxu Zhang; Xiang Zhou; Quanbin Zhang; Shaohui Jiang; Ruihao Wang; Guoqian Zhang; Huaiyu Lei; Shengqu Lin
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.