Literature DB >> 21506064

A comparison of functional and physical properties of self-expanding intracranial stents [Neuroform3, Wingspan, Solitaire, Leo+, Enterprise].

O Krischek1, E Miloslavski, S Fischer, S Shrivastava, H Henkes.   

Abstract

PURPOSE: 5 self-expanding intracranial stents Neuroform (N), Wingspan (W), Solitaire (S), Leo(+) (L), and Enterprise (E) were subjected to an in vitro examination and comparison of their physical features and functional properties in order to better understand the clinical advantages and potential limitations of each device.
MATERIAL AND METHODS: The following features were examined for each stent: visual appearance, radial strength, wall apposition, bending stiffness, gator backing, kink resistance, ovalization, vessel wall coverage, cell size, ease of delivery.
RESULTS: Given are rankings for the 5 stents: radial force at 50% oversizing: L<N<E<S<W; radial force at 15% oversizing L<E<S<N<W; wall apposition: E<N=W<S; bending stiffness: N<L<W<S<E; gator backing: N and W only; kink resistance: N=W<E<S<L; ovalization: W<L<S<N<E; vessel wall coverage: S<E<N<W<L; cell size: L<W<E<N<S; ease of delivery: W<N<L<E<S. A comparative analysis of the in vitro test results with the clinical experience of the authors is presented in this paper.
CONCLUSION: The 5 stents have fundamentally different features and there is no stent that is superior in all tested aspects. The selection in an individual treatment should be based on clinical and technical requirements. © Georg Thieme Verlag KG Stuttgart · New York.

Mesh:

Year:  2011        PMID: 21506064     DOI: 10.1055/s-0031-1271681

Source DB:  PubMed          Journal:  Minim Invasive Neurosurg        ISSN: 0946-7211


  67 in total

1.  LEO Baby Stent Use following Balloon-Assisted Coiling: Single- and Dual-Stent Technique--Immediate and Midterm Results of 29 Consecutive Patients.

Authors:  P Machi; V Costalat; K Lobotesis; C Ruiz; Y B Cheikh; O Eker; G Gascou; F Danière; C Riquelme; A Bonafé
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-13       Impact factor: 3.825

2.  Y-crossing of braided stents with stents and flow diverters does not cause significant stenosis in bench-top studies.

Authors:  Alina Makoyeva; Tim E Darsaut; Igor Salazkin; Jean Raymond
Journal:  Interv Neuroradiol       Date:  2013-12-18       Impact factor: 1.610

3.  In-vitro thrombogenicity assessment of flow diversion and aneurysm bridging devices.

Authors:  Gaurav Girdhar; Junwei Li; Larisa Kostousov; John Wainwright; Wayne L Chandler
Journal:  J Thromb Thrombolysis       Date:  2015-11       Impact factor: 2.300

4.  Comparison between Solitaire™ AB and Enterprise stent-assisted coiling for intracranial aneurysms.

Authors:  Hua-Wei Ye; Ya-Qi Liu; Qiu-Jing Wang; Tao Zheng; Xu-Bo Cui; Yu-Yuan Gao; Ling-Feng Lai; Xin Zhang; Xi-Feng Li; Shi-Xing Su; Xu-Ying He; Chuan-Zhi Duan
Journal:  Exp Ther Med       Date:  2015-05-08       Impact factor: 2.447

5.  Endovascular reconstruction of internal carotid artery dissection in patients with acute ischemic stroke using the Wingspan stent.

Authors:  Eduardo Murias Quintana; Pedro Vega Valdés; Edison Morales Deza; Alberto Gil Garcia; Hugo Cuellar Saenz; Alba L Salgado Bernal; María Cadenas Rodríguez; Lorena Benavente Fernández; Montserrat González Delgado; Nuria Riesco Pérez; Davinia Larrosa Campo; Sergio Calleja Puerta
Journal:  Interv Neuroradiol       Date:  2016-10-13       Impact factor: 1.610

6.  Solitaire salvage: a stent retriever-assisted catheter reduction technical report.

Authors:  Phillip Vaughan Parry; Alejandro Morales; Brian Thomas Jankowitz
Journal:  BMJ Case Rep       Date:  2015-10-28

7.  In Vitro Evaluation of Fusiform-Shaped Stents for Wide-Neck Intracranial Aneurysm Treatment.

Authors:  Zhen Yu Jia; Yuan Yuan Jiang; Jung Min Woo; Seon Moon Hwang; Ok Kyun Lim; Tae Il Kim; Jung Cheol Park; Hee Sun Lee; Eun Sang Kim; Deok Hee Lee
Journal:  Neurointervention       Date:  2018-08-31

8.  Super-selective coil embolization of a basilar perforator artery aneurysm previously treated by the stent-in-stent technique, using an extremely soft bare coil delivered through a one-marker microcatheter.

Authors:  Yves Chau; Marina Sachet; Jacques Sédat
Journal:  Interv Neuroradiol       Date:  2017-07-20       Impact factor: 1.610

9.  Disappearance of a small intracranial aneurysm as a result of vessel straightening and in-stent stenosis following use of an Enterprise vascular reconstruction device.

Authors:  Koichiro Takemoto; Satoshi Tateshima; Sachin Rastogi; Nestor Gonzalez; Reza Jahan; Gary Duckwiler; Fernando Vinuela
Journal:  BMJ Case Rep       Date:  2013-01-17

10.  Stent-Assisted Clip Placement for Complex Internal Carotid Artery Intracranial Aneurysms.

Authors:  Adnan I Qureshi; Morad Chughtai; Asif A Khan; M Fareed K Suri; Gregory T Sherr
Journal:  J Vasc Interv Neurol       Date:  2016-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.