Literature DB >> 21496657

An evaluation of root ZX and iPex apex locators: an in vivo study.

Eva K Stöber1, Fernando Duran-Sindreu, Montserrat Mercadé, Jorge Vera, Rufino Bueno, Miguel Roig.   

Abstract

INTRODUCTION: iPex is a fourth-generation apex locator (EAL) that has not yet been tested in vivo. The purpose of this study was to compare the accuracy of the Root ZX (J Morita Corp, Tokyo, Japan) and iPex (NSK, Tochigi, Japan) EALs.
METHODS: The working length (WL) was determined electronically for 40 root canals of human teeth with a K-file and one of the two EALs. The files were fixed at the WL, and the teeth were extracted. The apical 4 mm of each canal was trimmed to expose the file tip. The samples were observed under a scanning electron microscope, and the distance from the file tip to the point 0.5 mm coronal to the major foramen (the actual WL) was measured. The data were analyzed using the Mann-Whitney U test, and significance was set at P < .05.
RESULTS: No statistically significant differences were found between the Root ZX and iPex devices. The mean distance from the actual WL to the file tip was 0.146 ± 0.43 mm for the Root ZX and 0.128 ± 0.49 mm for the iPex. In determining the actual WL, the Root ZX was accurate 72% of the time to ± 0.5 mm and 100% of the time to ± 1 mm, whereas the iPex was accurate 57.8% of the time to ± 0.5 mm and 100% of the time to ± 1 mm.
CONCLUSIONS: Under the in vivo clinical conditions of this study, no statistically significant differences were observed between the Root ZX and iPex EALs.
Copyright © 2011 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21496657     DOI: 10.1016/j.joen.2011.02.002

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  10 in total

1.  Formulating a regression equation for determination of working length in primary molars using apex locators: a clinical study.

Authors:  A Dandempally; R Muppa; M K Duddu; P Bhupatiraju; S Nallanchakrava
Journal:  Eur Arch Paediatr Dent       Date:  2013-06-18

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  Evaluation of electrical impedance ratio measurements in accuracy of electronic apex locators.

Authors:  Pil-Jong Kim; Hong-Gee Kim; Byeong-Hoon Cho
Journal:  Restor Dent Endod       Date:  2014-12-26

4.  Evaluation of Dentaport ZX and Raypex 6 electronic apex locators: an in vivo study.

Authors:  S Moscoso; K Pineda; J Basilio; C Alvarado; M Roig; F Duran-Sindreu
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-03-01

5.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

6.  Comparative Evaluation of Accuracy of Ipex, Root Zx Mini, and Epex Pro Apex Locators in Teeth with Artificially Created Root Perforations in Presence of Various Intracanal Irrigants.

Authors:  Sakshi Bilaiya; Pallav Mahesh Patni; Pradeep Jain; Sanket Hans Pandey; Swadhin Raghuwanshi; Bhupesh Bagulkar
Journal:  Eur Endod J       Date:  2020-03-19

7.  Influence of calcium hydroxide residues after using different irrigants on the accuracy of two electronic apex locators: An in vitro study.

Authors:  Nooshin Sadat Shojaee; Zahra Zaeri; Mohammad Mehdi Shokouhi; Fereshteh Sobhnamayan; Alireza Adl
Journal:  Dent Res J (Isfahan)       Date:  2020-01-21

8.  An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method.

Authors:  Kalyan Vinayak Chakravarthy Pishipati
Journal:  Iran Endod J       Date:  2013-08-01

9.  An in vitro comparison of root canal length determination by DentaPort ZX and iPex apex locators.

Authors:  Nikhil Puri; Rupali Chadha; Pragya Kumar; Komal Puri
Journal:  J Conserv Dent       Date:  2013-11

Review 10.  Accuracy of different generations of apex locators in determining working length; a systematic review and meta-analysis.

Authors:  Kaveh Nasiri; Karl-Thomas Wrbas
Journal:  Saudi Dent J       Date:  2021-09-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.