Literature DB >> 21483877

Evaluation of the effect of retightening and mechanical cycling on preload maintenance of retention screws.

Juliana Aparecida Delben1, Erica Alves Gomes, Valentim Adelino Ricardo Barão, Wirley Gonçalves Assunção.   

Abstract

PURPOSE: The aim of this study was to evaluate the effect of retightening and mechanical cycling on the maintenance of preload in the retention screws of implant-supported crowns.
MATERIALS AND METHODS: Five groups (n = 12 in each group) of implant-crown assemblies were created according to different abutments and veneering materials: Gold UCLA abutments cast in gold alloy veneered with ceramic (group 1) and light-curing resin (group 2), UCLA abutments cast in titanium veneered with ceramic (group 3) and light-curing resin (group 4), and zirconia abutments with ceramic veneering (group 5). The crowns were attached to implants by gold retention screws. The assemblies were submitted to mechanical cycling for a total of 1 million cycles. Removal torque measurements were performed initially and after each period of 100,000 cycles; this was followed by screw retightening with 35 Ncm of torque. Data were evaluated with analysis of variance, the paired t test, and the Student t test.
RESULTS: All groups exhibited reduced removal torque values in comparison to insertion torque initially and after all periods of mechanical cycling. Analysis of variance revealed statistically significant differences among the groups, but there were no differences among the periods of mechanical cycling or for the interaction between groups and cycling. Group 1 was the only group that did not show a difference between removal torque values of different cycling periods. The Student t test did not reveal statistically significant differences between mean removal torque values obtained before and after mechanical cycling, except for group 4.
CONCLUSIONS: Torque maintenance allowed screw joint stability in the present study. Mechanical cycling associated with the retightening of abutment screws did not influence the maintenance of insertion torque.

Mesh:

Substances:

Year:  2011        PMID: 21483877

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  10 in total

1.  Effect of cyclic loading and retightening on reverse torque value in external and internal implants.

Authors:  Woong-Rae Cho; Yoon-Hyuk Huh; Chan-Jin Park; Lee-Ra Cho
Journal:  J Adv Prosthodont       Date:  2015-08-18       Impact factor: 1.904

2.  Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants.

Authors:  Hnd Hadi Mohammed; Jin-Han Lee; Ji-Myung Bae; Hye-Won Cho
Journal:  J Adv Prosthodont       Date:  2016-02-23       Impact factor: 1.904

3.  Effect of dynamic cyclic loading on screw loosening of retightened versus new abutment screw in both narrow and standard implants (in-vitro study).

Authors:  Eman Mohammed Nasr Attiah; Attiah Ali AlGendy; Tamer Mohamed Nasr Mostafa
Journal:  Int J Implant Dent       Date:  2020-07-28

4.  Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading.

Authors:  Mai Ahmed Yousry El-Sheikh; Tamer Mohamed Nasr Mostafa; Mohamed Maamoun El-Sheikh
Journal:  Int J Implant Dent       Date:  2018-12-03

5.  Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force.

Authors:  Yuseung Yi; Seong-Joo Heo; Jai-Young Koak; Seong-Kyun Kim
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

Review 6.  Review of the Mechanical Behavior of Different Implant-Abutment Connections.

Authors:  Ana Sofia Vinhas; Carlos Aroso; Filomena Salazar; Paula López-Jarana; José Vicente Ríos-Santos; Mariano Herrero-Climent
Journal:  Int J Environ Res Public Health       Date:  2020-11-23       Impact factor: 3.390

7.  The effect of the digital manufacturing technique of cantilevered implant-supported frameworks on abutment screw preload.

Authors:  Shahad Mohammmed Altuwaijri; Hanan Nejer Alotaibi; Talal Mughaileth Alnassar
Journal:  J Adv Prosthodont       Date:  2022-02-25       Impact factor: 1.904

8.  Analysis of Torque Maintenance and Fracture Resistance after Fatigue in Retention Screws Made of Different Metals for Screw-Retained Implant-Borne Prosthesis Joints.

Authors:  Maria Beatriz Bello Taborda; Gabriela Sumie Yaguinuma Gonçalves; Cecília Alves de Sousa; Wirley Gonçalves Assunção
Journal:  Int J Dent       Date:  2021-11-18

9.  Healing Abutment Distortion in Implant Prostheses: An In Vitro Study.

Authors:  Chandrashekhar Pandey; Bishwa Prakash Bhattarai; Apiwat Riddhabhaya; Natthamet Wongsirichat; Dinesh Rokaya
Journal:  J Funct Biomater       Date:  2022-06-21

10.  Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading.

Authors:  Behnaz Ebadian; Amirhossein Fathi; Saba Khodadad
Journal:  Int J Dent       Date:  2021-07-19
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.