Literature DB >> 21482439

Long-term clinical and angiographic outcomes of patients with sirolimus-eluting stent fracture.

Ung Kim1, Doo-Il Kim, Dong-Kie Kim, Sang-Hoon Seol, Jae-Sik Jang, Tae-Hyun Yang, Dae-Kyeong Kim, Dong-Soo Kim, Ho-Ki Min, Kang-Joo Choi.   

Abstract

BACKGROUND: There is limited data on the long-term outcomes of patients with sirolimus-eluting stent (SES) fracture.
METHODS: After performing index percutaneous coronary intervention (PCI) with SES from Dec. 2003 to Dec. 2007, the patients with stent fracture (SF) detected by angiography at follow-up or adverse events were enrolled. SF was angiographically defined as complete or partial separation of the stent and this had been contiguous after stenting. The patients were divided into two groups as SF without in-stent restenosis (ISR) (Group 1, n=55) and SF with ISR (Group 2, n=44) and the patients were clinically and angiographically followed up after the detection of SF. The major adverse cardiovascular events (MACE), including cardiac death, myocardial infarction (MI), target vessel revascularization (TVR), and stent thrombosis (ST) were determined.
RESULTS: At SF detection, most patients were asymptomatic (92.7% in Group 1 vs. 43.2% in Group 2, p=0.001). Acute coronary syndrome developed in 18.2% of Group 2, but it developed in only 1.8% of the patients of Group 1. Clinical follow-up was done in all patients for 675 ± 375 days from SF detection. From the detection of SF, MACEs were noted in 32.3% of the patients (10.9% in Group 1 vs. 59.1% in Group 2, p=0.001), mostly from TVR (9.1% in Group 1 vs. 54.5% in Group 2, p=0.001).
CONCLUSION: The long-term follow-up results of SF after SES implantation show an increased risk of adverse cardiac events, and especially for the patients with ISR.
Copyright © 2011. Published by Elsevier Ireland Ltd.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21482439     DOI: 10.1016/j.ijcard.2011.01.013

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164


  4 in total

1.  Prevalence and prognosis of coronary stent gap detected by multi-detector CT: a follow-up study.

Authors:  Xinghua Zhang; Li Yang; Haiyue Ju; Fan Zhang; Jian Wu; Bin He; Yundai Chen
Journal:  Eur Radiol       Date:  2012-04-27       Impact factor: 5.315

2.  Impact of sirolimus-eluting stent fractures without early cardiac events on long-term clinical outcomes: a multislice computed tomography study.

Authors:  Tsuyoshi Ito; Masashi Kimura; Mariko Ehara; Mitsuyasu Terashima; Kenya Nasu; Yoshihisa Kinoshita; Maoto Habara; Etsuo Tsuchikane; Takahiko Suzuki
Journal:  Eur Radiol       Date:  2014-02-25       Impact factor: 5.315

3.  Incidence of coronary drug-eluting stent fracture: A systematic review and meta-analysis.

Authors:  Yang Chen; Dandan Li; Yanhui Liao; Xiongda Yao; Yuehua Ruan; Kai Zou; Hanhui Liao; Jingwen Ding; Hao Qin; Zuozhong Yu; Yuanbin Zhao; Longlong Hu; Renqiang Yang
Journal:  Front Cardiovasc Med       Date:  2022-08-23

4.  Optical Coherence Tomography-guided Treatment of Multiple Stent Fracture in Patients with Overlapping Drug-eluting Stents.

Authors:  Jung Hee Lee; Ung Kim
Journal:  Korean Circ J       Date:  2018-11       Impact factor: 3.243

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.