| Literature DB >> 21479643 |
G J M Tuijthof1, M M de Vaal, I N Sierevelt, L Blankevoort, M P J van der List.
Abstract
PURPOSE: During arthroscopies, bleeding episodes occur as a result of tissue damage. Irrigation systems assist in minimizing these disturbances. The performance of three arthroscopic irrigation systems in clearing bleeding episodes was evaluated objectively.Entities:
Mesh:
Year: 2011 PMID: 21479643 PMCID: PMC3195766 DOI: 10.1007/s00167-011-1495-z
Source DB: PubMed Journal: Knee Surg Sports Traumatol Arthrosc ISSN: 0942-2056 Impact factor: 4.342
Fig. 1Three arthroscopic irrigation systems were assessed for their performance with automated blood detection: a gravity infusion, b single roller pump, and c double roller pump. For the gravity infusion, two bags were routinely hung on a pole at a standard height, but only one was used at the same time
Ninety-nine shoulder arthroscopies were matched for type of surgery, instrument use, and operation time
| Gravity infusion | Single roller pump | Double roller pump | ||
|---|---|---|---|---|
| SLAP (superior labrum anterior posterior) lesion repair | Number of procedures [#] | 2 | 2 | 2 |
| Mean (SD) OT [min.:sec.] | 29:27 (9:17) | 37:38 (6:31) | 25:19 (7:53) | |
| Mean (SD) IU [% of OT] | 5.5 (5.0) | 2.0 (0.0) | 5.5 (6.4) | |
| Acromioplasty and rotator cuff tear repair | Number of procedures [#] | 1 | 1 | 1 |
| Mean (SD) OT [min.:sec.] | 33:16 | 47:53 | 29:20 | |
| Mean (SD) IU [% of OT] | 55.0 | 32.0 | 62.0 | |
| Acromioplasty | Number of procedures [#] | 14 | 14 | 14 |
| Mean (SD) OT [min.:sec.] | 36:52 (13:58) | 28:51 (7:24) | 32:12 (7:37) | |
| Mean (SD) IU [% of OT] | 64.2 (13.7) | 70.2 (9.4) | 68.8 (8.4) | |
| Release frozen shoulder | Number of procedures [#] | 4 | 4 | 4 |
| Mean (SD) OT [min.:sec.] | 13:29 (4:38) | 12:28 (4:14) | 15:28 (8:22) | |
| Mean (SD) IU [% of OT] | 40.0 (18.1) | 45.8 (19.5) | 50.5 (12.4) | |
| Rotator cuff tear repair | Number of procedures [#] | 5 | 5 | 5 |
| Mean (SD) OT [min.:sec.] | 42:06 (17:44) | 49:24 (23:44) | 37:44 (20:27) | |
| Mean (SD) IU [% of OT] | 6.6 (11.0) | 14.8 (13.6) | 14.4 (18.1) | |
| Capsule repair | Number of procedures [#] | 7 | 7 | 7 |
| Mean (SD) OT [min.:sec.] | 26:28 (16:00) | 26:44 (16:30) | 21:23 (8:13) | |
| Mean (SD) IU [% of OT] | 0.4 (1.1) | 0.9 (2.3) | 3.9 (6.3) |
OT Operation time, IU instrument usage
Results of the four outcome measures for all operations together and for the two subgroups removal and reconstruction
| Gravity infusion | Single roller pump | Double roller pump | ||
|---|---|---|---|---|
| All operations | Median (min.–max.) bleeding episode [% of OT] | 6.6 (0.0–43.6) | 3.7 (0.2–46.4) | 3.3 (0.0–19.3) |
| Median (min.–max.) ATB [sec.] | 75.0 (0.0–927.7) | 65.3 (5.7–457.1) | 54.8 (0.0–339.6) | |
| Median (min.–max.) of MCRB [sec.] | 4.8 (0.0–20.3) | 4.3 (0.0–9.8) | 3.7 (0.0–7.5) | |
| Subgroup removal procedures | Median (min.–max.) ATB [sec.] | 230.0 (34.3–927.7)* | 116.1 (31.2–457.1)* | 88.2 (42.0–281.5)* |
| Median (min.–max.) of MCRB [sec.] | 5.4 (2.9–7.2) | 4.2 (2.9–8.2) | 4.3 (3.2–7.5) | |
| Subgroup reconstruction procedures | Median (min.–max.) ATB [sec.] | 6.5 (0.0–268.9) | 12.3 (5.7–324.5) | 4.6 (0.0–137.2) |
| Median (min.–max.) of MCRB [sec.] | 0.0 (0.0–7.4) | 4.0 (0.0–6.6)** | 0.0 (0.0–4.2)** |
If no symbols are added to the values, no significant differences were present. For ‘*’, a significant difference (MWU, P < 0.05) was present between the gravity infusion and both types of roller pumps, whereas no significant difference was present between the roller pumps. For ‘**’, a significant difference (MWU, P < 0.05) was present between the single roller and the double roller pump solely. OT Operation time, ATB absolute total bleeding episode time per operation, MCRB mean clinically relevant bleeding episode time per operation
Fig. 2Result of automated detection of bleeding episodes of one arthroscopic frozen shoulder release operation. The percentage of red-colored arthroscopic image area per video frame is automatically detected and plotted against time (gray line). The threshold level of 20% is indicated by the striped horizontal line. All frames that demonstrate more than 20% red image area coverage are included to determine the presence of bleeding episodes. One bleeding episode is indicated by the vertical black lines
Fig. 3Boxplot is shown of the bleeding episode time as percentage of the operation time. The results are shown for the subgroups removal procedures consisting of 18 matched operations and reconstruction procedures consisting of 12 matched operations (Tables 1, 2)