Literature DB >> 21458652

Retentive force and magnetic flux leakage of magnetic attachment in various keeper and magnetic assembly combinations.

Mikage Hasegawa1, Yoshitada Umekawa, Eiich Nagai, Tomohiko Ishigami.   

Abstract

STATEMENT OF PROBLEM: Magnetic attachments are commonly used for overdentures. However, it can be difficult to identify and provide the same type and size of magnetic assembly and keeper if a repair becomes necessary. Therefore, the size and type may not match.
PURPOSE: This study evaluated the retentive force and magnetic flux strength and leakage of magnetic attachments in different combinations of keepers and magnetic assemblies.
MATERIAL AND METHODS: For 6 magnet-keeper combinations using 4 sizes of magnets (GIGAUSS D400, D600, D800, and D1000) (n=5), retentive force was measured 5 times at a crosshead speed of 5 mm/min in a universal testing machine. Magnetic flux strength was measured using a Hall Effect Gaussmeter. Data were statistically analyzed using a 1-way ANOVA, and between-group differences were analyzed with Tukey's HSD post hoc test (α=.05).
RESULTS: The mean retentive force of the same-size magnet-keeper combinations was 3.2 N for GIGAUSS D400 and 5.1 N for GIGAUSS D600, but was significantly reduced when using larger magnets (P<.05). Magnetic flux leakage was significantly lower for corresponding size combinations.
CONCLUSIONS: Size differences influence the retentive force and magnetic flux strength of magnetic attachments. Retentive force decreased due to the closed field structure becoming incomplete and due to magnetic field leakage.
Copyright © 2011 The Editorial Council of the Journal of Prosthetic Dentistry. Published by Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21458652     DOI: 10.1016/S0022-3913(11)60042-5

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  5 in total

1.  Radiofrequency heating and magnetically induced displacement of dental magnetic attachments during 3.0 T MRI.

Authors:  K Miyata; M Hasegawa; Y Abe; T Tabuchi; T Namiki; T Ishigami
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

2.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

3.  Magnetization and demagnetization of magnetic dental attachments in a 3-T MRI system.

Authors:  Norio Hayashi; Akio Ogura; Toshio Tsuchihashi; Daisuke Takahashi; Tsuyoshi Matsuda; Shinya Seino; Tsukasa Doi
Journal:  Radiol Phys Technol       Date:  2017-04-27

Review 4.  Factors Affecting the Attractive Force of Dental Magnetic Attachment: A Literature Review for Guiding Dentists in Clinical Application.

Authors:  An-Nissa Kusumadewi; Lisda Damayanti
Journal:  Int J Dent       Date:  2022-06-14

5.  Novel Magnetic Attachment System Manufactured Using High-Frequency Heat Treatment and Stamp Technique: Introduction and Basic Performance.

Authors:  Adityakrisna Yoshi Putra Wigianto; Yuichi Ishida; Takashi Matsuda; Takaharu Goto; Megumi Watanabe; Tetsuo Ichikawa
Journal:  Dent J (Basel)       Date:  2022-05-02
  5 in total