Ipsilateral pulmonary vein isolation performed by a single continuous circular lesion: role of pulmonary vein mapping during ablation.

Alexander Fürnkranz1, Julian K R Chun Julian, Boris Schmidt, Peter Wohlmuth, Roland Tilz, Karl-Heinz Kuck, Feifan Ouyang.   

Abstract

AIMS: Complete pulmonary vein isolation (CPVI) can be achieved by continuous circular lesions (CCL) around the ipsilateral pulmonary veins (PVs) guided by a 3D-mapping system. We investigated whether CPVI can be achieved with a single CCL around the isplilateral PVs without recording PV activity during ablation. METHODS AND
RESULTS: Fifty patients with atrial fibrillation underwent ablation of CCLs around ipsilateral PVs guided by 3D mapping. One or two Lasso catheters were placed within the PVs. Lasso tracings were hidden to physicians during ablation. After completion of CCLs, Lasso tracings were evaluated. If PV activation was present, conduction gaps (CGs) were identified and ablated with guidance by the local electrogram and the Lasso catheter(s). In 21 patients (42%), CPVI was achieved after ablation of a single CCL around ipsilateral PVs. Pulmonary vein isolation was achieved in 43 patients (86%) in the right-sided PVs and in 21 patients (42%) in the left-sided PVs. In the remaining patients, there were eight CGs in right-sided CCLs and 40 CGs in left-sided CCLs. Conduction gaps along the left CCLs were found at the ridge between the PV ostia and the left atrial appendage in 27 out of 40 CGs (68%). Mean time from the P-wave onset to the earliest PV potential was 112±35 ms in the presence of a CG at the roof, and 166±59 ms in patients with CGs at other locations in left-sided CCLs (P<0.05).
CONCLUSION: Complete pulmonary vein isolation is difficult to achieve with a single CCL around ipsilateral PVs without continuous recording of PV activation during ablation.

Entities:  

Mesh:

Year:  2011        PMID: 21454334     DOI: 10.1093/europace/eur067

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  7 in total

1.  General anesthesia improves contact force and reduces gap formation in pulmonary vein isolation: a comparison with conscious sedation.

Authors:  Akio Chikata; Takeshi Kato; Takanori Yaegashi; Satoru Sakagami; Chieko Kato; Takahiro Saeki; Keiichi Kawai; Shin-Ichiro Takashima; Hisayoshi Murai; Soichiro Usui; Hiroshi Furusho; Shuichi Kaneko; Masayuki Takamura
Journal:  Heart Vessels       Date:  2017-03-04       Impact factor: 2.037

2.  THERMOCOOL® SMARTTOUCH® CATHETER - The Evidence So Far for Contact Force Technology and the Role of VISITAG™ MODULE.

Authors:  Tina Lin; Feifan Ouyang; Karl-Heinz Kuck; Roland Tilz
Journal:  Arrhythm Electrophysiol Rev       Date:  2014-05-30

3.  Remote magnetic navigation for circumferential pulmonary vein ablation: single-catheter technique or additional use of a circular mapping catheter?

Authors:  Dirk Vollmann; Lars Lüthje; Joachim Seegers; Christian Sohns; Samuel Sossalla; Jan Sohns; Christian Röver; Gerd Hasenfuß; Markus Zabel
Journal:  J Interv Card Electrophysiol       Date:  2014-06-04       Impact factor: 1.900

4.  Comparison of contact force-guided procedure with non-contact force-guided procedure during left atrial mapping and pulmonary vein isolation: impact of contact force on recurrence of atrial fibrillation.

Authors:  Hisaki Makimoto; Christian-H Heeger; Tina Lin; Andreas Rillig; Andreas Metzner; Erik Wissner; Shibu Mathew; Sebastian Deiss; Peter Rausch; Christine Lemeš; Karl-Heinz Kuck; Feifan Ouyang; Roland Richard Tilz
Journal:  Clin Res Cardiol       Date:  2015-04-18       Impact factor: 5.460

Review 5.  Atrial Fibrillation in Congenital Heart Disease.

Authors:  Irene Martín de Miguel; Pablo Ávila
Journal:  Eur Cardiol       Date:  2021-03-09

6.  Optimal Force-Time Integral for Pulmonary Vein Isolation According to Anatomical Wall Thickness Under the Ablation Line.

Authors:  Akio Chikata; Takeshi Kato; Satoru Sakagami; Chieko Kato; Takahiro Saeki; Keiichi Kawai; Shin-Ichiro Takashima; Hisayoshi Murai; Soichiro Usui; Hiroshi Furusho; Shuichi Kaneko; Masayuki Takamura
Journal:  J Am Heart Assoc       Date:  2016-03-15       Impact factor: 5.501

7.  Novel Clue to Locate Conduction Gaps in the Pulmonary Vein Isolation Ablation Line.

Authors:  Hai-Yang Xie; Xiao-Gang Guo; Jian-du Yang; Yan-Qiao Chen; Zhong-Jing Cao; Qi Sun; Jian Ma
Journal:  Front Cardiovasc Med       Date:  2021-07-12