Four-dimensional cone-beam computed tomography and digital tomosynthesis reconstructions using respiratory signals extracted from transcutaneously inserted metal markers for liver SBRT.

Justin C Park1, Sung Ho Park, Jong Hoon Kim, Sang Min Yoon, Su Ssan Kim, Jin Sung Kim, Zhaowei Liu, Tyler Watkins, William Y Song.   

Abstract

PURPOSE: Respiration-induced intrafraction target motion is a concern in liver cancer radiotherapy, especially in stereotactic body radiotherapy (SBRT), and therefore, verification of its motion is necessary. An effective means to localize the liver cancer is to insert metal fiducial markers to or near the tumor with simultaneous imaging using cone-beam computed tomography (CBCT). Utilizing the fiducial markers, the authors have demonstrated a method to generate breath-induced motion signal of liver for reconstructing 4D digital tomosynthesis (4DDTS) and 4DCBCT images based on phasewise and/or amplitudewise sorting of projection data.
METHODS: The marker extraction algorithm is based on template matching of a prior known marker image and has been coded to optimally extract marker positions in CBCT projections from the On-Board Imager (Varian Medical Systems, Palo Alto, CA). To validate the algorithm, multiple projection images of moving thorax phantom and five patient cases were examined. Upon extraction of the motion signals from the markers, 4D image sorting and image reconstructions were subsequently performed. In the case of incomplete signals due to projections with missing markers, the authors have implemented signal profiling to replace the missing portion.
RESULTS: The proposed marker extraction algorithm was shown to be very robust and accurate in the phantom and patient cases examined. The maximum discrepancy of the algorithm predicted marker location versus operator selected location was < 1.2 mm, with the overall average of 0.51 +/- 0.15 mm, for 500 projections. The resulting 4DDTS and 4DCBCT images showed clear reduction in motion-induced blur of the markers and the anatomy for an effective image guidance. The signal profiling method was useful in replacing missing signals.
CONCLUSIONS: The authors have successfully demonstrated that motion tracking of fiducial markers and the subsequent 4D reconstruction of CBCT and DTS are possible. Due to the significant reduction in motion-induced image blur, it is anticipated that such technology will be useful in image-guided liver SBRT treatments.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21452740     DOI: 10.1118/1.3544369

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  A novel technique for markerless, self-sorted 4D-CBCT: feasibility study.

Authors:  Irina Vergalasova; Jing Cai; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-03       Impact factor: 4.071

2.  Extraction of tumor motion trajectories using PICCS-4DCBCT: a validation study.

Authors:  Zhihua Qi; Guang-Hong Chen
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

3.  A hybrid reconstruction algorithm for fast and accurate 4D cone-beam CT imaging.

Authors:  Hao Yan; Xin Zhen; Michael Folkerts; Yongbao Li; Tinsu Pan; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

4.  A method for volumetric imaging in radiotherapy using single x-ray projection.

Authors:  Yuan Xu; Hao Yan; Luo Ouyang; Jing Wang; Linghong Zhou; Laura Cervino; Steve B Jiang; Xun Jia
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

5.  Multiple anatomy optimization of accumulated dose.

Authors:  W Tyler Watkins; Joseph A Moore; James Gordon; Geoffrey D Hugo; Jeffrey V Siebers
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

6.  Preliminary clinical evaluation of a 4D-CBCT estimation technique using prior information and limited-angle projections.

Authors:  You Zhang; Fang-Fang Yin; Tinsu Pan; Irina Vergalasova; Lei Ren
Journal:  Radiother Oncol       Date:  2015-03-26       Impact factor: 6.280

7.  Dosimetric feasibility of real-time MRI-guided proton therapy.

Authors:  M Moteabbed; J Schuemann; H Paganetti
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

8.  A Novel Method of Cone Beam CT Projection Binning Based on Image Registration.

Authors:  Seonyeong Park; Siyong Kim; Byongyong Yi; Geoffrey Hugo; H Michael Gach; Yuichi Motai
Journal:  IEEE Trans Med Imaging       Date:  2017-03-31       Impact factor: 10.048

9.  Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition.

Authors:  Andrew P Santoso; Kwang H Song; Yujiao Qin; Stephen J Gardner; Chang Liu; Indrin J Chetty; Benjamin Movsas; Munther Ajlouni; Ning Wen
Journal:  Radiat Oncol       Date:  2016-07-29       Impact factor: 3.481

10.  Four-dimensional measurement of the displacement of metal clips or postoperative surgical staples during 320-multislice computed tomography scanning of gastric cancer.

Authors:  Hideomi Yamashita; Kae Okuma; Wataru Takahashi; Akira Sakumi; Akihiro Haga; Kenji Ino; Masaaki Akahane; Kuni Ohtomo; Keiichi Nakagawa
Journal:  Radiat Oncol       Date:  2012-08-10       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.