Literature DB >> 21444968

Non-invasive continuous core temperature measurement by zero heat flux.

L P J Teunissen1, J Klewer, A de Haan, J J de Koning, H A M Daanen.   

Abstract

Reliable continuous core temperature measurement is of major importance for monitoring patients. The zero heat flux method (ZHF) can potentially fulfil the requirements of non-invasiveness, reliability and short delay time that current measurement methods lack. The purpose of this study was to determine the performance of a new ZHF device on the forehead regarding these issues. Seven healthy subjects performed a protocol of 10 min rest, 30 min submaximal exercise (average temperature increase about 1.5 °C) and 10 min passive recovery in ambient conditions of 35 °C and 50% relative humidity. ZHF temperature (T(zhf)) was compared to oesophageal (T(es)) and rectal (T(re)) temperature. ΔT(zhf)-T(es) had an average bias ± standard deviation of 0.17 ± 0.19 °C in rest, -0.05 ± 0.18 °C during exercise and -0.01 ± 0.20 °C during recovery, the latter two being not significant. The 95% limits of agreement ranged from -0.40 to 0.40 °C and T(zhf) had hardly any delay compared to T(es). T(re) showed a substantial delay and deviation from T(es) when core temperature changed rapidly. Results indicate that the studied ZHF sensor tracks T(es) very well in hot and stable ambient conditions and may be a promising alternative for reliable non-invasive continuous core temperature measurement in hospital.

Entities:  

Mesh:

Year:  2011        PMID: 21444968     DOI: 10.1088/0967-3334/32/5/005

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  15 in total

1.  Prediction of human core body temperature using non-invasive measurement methods.

Authors:  Reto Niedermann; Eva Wyss; Simon Annaheim; Agnes Psikuta; Sarah Davey; René Michel Rossi
Journal:  Int J Biometeorol       Date:  2013-06-13       Impact factor: 3.787

2.  Relationship between core temperature, skin temperature, and heat flux during exercise in heat.

Authors:  Xiaojiang Xu; Anthony J Karis; Mark J Buller; William R Santee
Journal:  Eur J Appl Physiol       Date:  2013-06-18       Impact factor: 3.078

3.  An IR Sensor Based Smart System to Approximate Core Body Temperature.

Authors:  Partha Pratim Ray
Journal:  J Med Syst       Date:  2017-07-10       Impact factor: 4.460

4.  Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting.

Authors:  James M Jack; Helen Ellicott; Christopher I Jones; Stephen A Bremner; Ian Densham; C Mark Harper
Journal:  J Clin Monit Comput       Date:  2019-01-08       Impact factor: 2.502

Review 5.  Smart health monitoring systems: an overview of design and modeling.

Authors:  Mirza Mansoor Baig; Hamid Gholamhosseini
Journal:  J Med Syst       Date:  2013-01-15       Impact factor: 4.460

6.  Umbilical temperature correlation with core and skin temperatures at rest, in the heat and during physical activity.

Authors:  Raymond J Roberge; Jung-Hyun Kim; Patrick Yorio; Aitor Coca; Yongsuk Seo; Tyler Quinn; Ali Aljaroudi; Jeffrey B Powell
Journal:  Int J Hyperthermia       Date:  2017-04-19       Impact factor: 3.914

7.  A Novel Non-Invasive Thermometer for Continuous Core Body Temperature: Comparison with Tympanic Temperature in an Acute Stroke Clinical Setting.

Authors:  Miloš Ajčević; Alex Buoite Stella; Giovanni Furlanis; Paola Caruso; Marcello Naccarato; Agostino Accardo; Paolo Manganotti
Journal:  Sensors (Basel)       Date:  2022-06-23       Impact factor: 3.847

8.  Theoretical and Experimental Studies of Epidermal Heat Flux Sensors for Measurements of Core Body Temperature.

Authors:  Yihui Zhang; Richard Chad Webb; Hongying Luo; Yeguang Xue; Jonas Kurniawan; Nam Heon Cho; Siddharth Krishnan; Yuhang Li; Yonggang Huang; John A Rogers
Journal:  Adv Healthc Mater       Date:  2015-05-07       Impact factor: 9.933

9.  Comparison of a Continuous Noninvasive Temperature to Monitor Core Temperature Measures During Targeted Temperature Management.

Authors:  Melissa Wagner; Krista Lim-Hing; Mary Ann Bautista; Brigid Blaber; Taghi Ryder; Joseph Haymore; Neeraj Badjatia
Journal:  Neurocrit Care       Date:  2021-04       Impact factor: 3.210

10.  Zero-Heat-Flux Thermometry for Non-Invasive Measurement of Core Body Temperature in Pigs.

Authors:  Maria Guschlbauer; Alexandra C Maul; Xiaowei Yan; Holger Herff; Thorsten Annecke; Anja Sterner-Kock; Bernd W Böttiger; Daniel C Schroeder