Experimental comparison of mesenteric vessel sealing and thermal damage between one bipolar and two ultrasonic shears devices.

E J Noble1, N J Smart, C Challand, K Sleigh, A Oriolowo, K B Hosie.   

Abstract

BACKGROUND: Several instruments are available for open and laparoscopic dissection, including electrothermal bipolar vessel sealers and ultrasonically coagulating shears. The vessel sealing ability of three devices in colorectal specimens was compared in an experimental study.
METHODS: Surgical specimens from patients scheduled for elective open or laparoscopic colorectal resection were allocated to one of the three devices. After removal of the surgical specimen, up to eight mesenteric vessels were dissected ex vivo and sealed using the allocated instrument. The vessel seal was tested for the maximum pressure at which it leaked and then assessed by a pathologist for depth of thermal tissue damage.
RESULTS: A total of 93 vessels from 18 patients were assessed ex vivo (LOTUS™ n = 33; Harmonic Ace® n = 30; LigaSure™ n = 30), a median of 6 (range 1-8) vessels per surgical specimen with a mean(s.d.) diameter of 1·06(0·70) mm and wall thickness of 0·29(0·19) mm. Mean(s.d.) bursting pressures were 1170(440), 1470(670) and 1510(740) mmHg with LOTUS(™) , Harmonic Ace® and LigaSure™ respectively. ANCOVA showed no difference in bursting pressure between the instruments (P = 0·058). The depth of thermal damage was significantly greater with LigaSure™ (3·37(1·44) mm) than with LOTUS(™) (2·18(0·99) mm; P < 0·001) or Harmonic Ace® (1·95(0·92) mm; P < 0·001).
CONCLUSION: All three instruments were equally good at sealing blood vessels, with bursting pressures well above physiological blood pressure levels. REGISTRATION NUMBER: NCT01121614 (http://www.clinicaltrials.gov).
Copyright © 2011 British Journal of Surgery Society Ltd. Published by John Wiley & Sons, Ltd.

Entities:  

Mesh:

Year:  2011        PMID: 21442611     DOI: 10.1002/bjs.7433

Source DB:  PubMed          Journal:  Br J Surg        ISSN: 0007-1323            Impact factor:   6.939


  7 in total

Review 1.  Effectiveness of electrothermal bipolar vessel-sealing devices versus other electrothermal and ultrasonic devices for abdominal surgical hemostasis: a systematic review.

Authors:  Petra F Janssen; Hans A M Brölmann; Judith A F Huirne
Journal:  Surg Endosc       Date:  2012-04-27       Impact factor: 4.584

2.  Prospective analysis of the sealing ability of the ENSEAL® G2 Articulating Tissue Sealer and transector on human mesenteric vessels in colorectal surgery.

Authors:  B Balachandran; G Melich; T Mustafa; S J Marecik; L M Prasad; M Gonzalez; S Sulo; F Dabbous; J J Park
Journal:  Tech Coloproctol       Date:  2017-01-31       Impact factor: 3.781

Review 3.  [Bleeding complications in bariatric surgery: Prophylaxis and therapy].

Authors:  H Spieker; A Dietrich
Journal:  Chirurg       Date:  2015-09       Impact factor: 0.955

4.  Postoperative efficacy and safety of vessel sealing: an experimental study on carotid arteries of the pig.

Authors:  Stéphane V Berdah; Christiaan Hoff; Peiman Hossein Poornoroozy; Peter Razek; Yves Van Nieuwenhove
Journal:  Surg Endosc       Date:  2012-02-21       Impact factor: 4.584

5.  Tissue effects in vessel sealing and transection from an ultrasonic device with more intelligent control of energy delivery.

Authors:  Duan Broughton; Alissa L Welling; Emily H Monroe; Kristen Pirozzi; John B Schulte; Jeffrey W Clymer
Journal:  Med Devices (Auckl)       Date:  2013-09-16

6.  Major vessel sealing in laparoscopic surgery for colorectal cancer: a single-center experience with 759 patients.

Authors:  Michele Grieco; Daniela Apa; Domenico Spoletini; Emanuela Grattarola; Massimo Carlini
Journal:  World J Surg Oncol       Date:  2018-06-01       Impact factor: 2.754

7.  Safety and effectiveness of a novel generator algorithm for bipolar vessel sealing: a randomised controlled chronic animal study.

Authors:  Bernhard Kraemer; Christos Tsaousidis; Stephan Kruck; Martin Schenk; Marcus Scharpf; Stefan Kommoss; Sara Brucker; Daniela Nuessle; Markus D Enderle; Ulrich Biber
Journal:  BMC Surg       Date:  2019-11-05       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.