| Literature DB >> 21424376 |
M B Kakakhel1, E S Baveas, A L Fielding, T Kairn, J Kenny, J V Trapp.
Abstract
The purpose of this work is to validate and automate the use of DYNJAWS; a new component module (CM) in the BEAMnrc Monte Carlo (MC) user code. The DYNJAWS CM simulates dynamic wedges and can be used in three modes; dynamic, step-and-shoot and static. The step-and-shoot and dynamic modes require an additional input file defining the positions of the jaw that constitutes the dynamic wedge, at regular intervals during its motion. A method for automating the generation of the input file is presented which will allow for the more efficient use of the DYNJAWS CM. Wedged profiles have been measured and simulated for 6 and 10 MV photons at three field sizes (5 cm × 5 cm, 10 cm × 10 cm and 20 cm × 20 cm), four wedge angles (15°, 30°, 45° and 60°), at d (max) and at 10 cm depth. Results of this study show agreement between the measured and the MC profiles to within 3% of absolute dose or 3 mm distance to agreement for all wedge angles at both energies and depths. The gamma analysis suggests that dynamic mode is more accurate than the step-and-shoot mode. The DYNJAWS CM is an important addition to the BEAMnrc code and will enable the MC verification of patient treatments involving dynamic wedges.Entities:
Mesh:
Year: 2011 PMID: 21424376 DOI: 10.1007/s13246-011-0060-x
Source DB: PubMed Journal: Australas Phys Eng Sci Med ISSN: 0158-9938 Impact factor: 1.430