Literature DB >> 21402488

Helmholtz-pair transmit coil with integrated receive array for high-resolution MRI of trabecular bone in the distal tibia at 7T.

Alexander C Wright1, Rostislav Lemdiasov, Thomas J Connick, Yusuf A Bhagat, Jeremy F Magland, Hee Kwon Song, Steven P Toddes, Reinhold Ludwig, Felix W Wehrli.   

Abstract

A Helmholtz-pair local transmit RF coil with an integrated four-element receive array RF coil and foot immobilization platform was designed and constructed for imaging the distal tibia in a whole-body 7T MRI scanner. Simulations and measurements of the B(1) field distribution of the transmit coil are described, along with SAR considerations for operation at 7T. Results of imaging the trabecular bone of three volunteers at 1.5T, 3T and 7T are presented, using identical 1.5T and 3T versions of the 7T four-element receive array. The spatially registered images reveal improved visibility for individual trabeculae and show average gains in SNR of 2.8× and 4.9× for imaging at 7T compared to 3T and 1.5T, respectively. The results thus display an approximately linear dependence of SNR with field strength and enable the practical utility of 7T scanners for micro-MRI of trabecular bone.
Copyright © 2011 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21402488      PMCID: PMC3085966          DOI: 10.1016/j.jmr.2011.02.019

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  33 in total

1.  Mechanical consequence of trabecular bone loss and its treatment: a three-dimensional model simulation.

Authors:  X E Guo; C H Kim
Journal:  Bone       Date:  2002-02       Impact factor: 4.398

2.  Musculoskeletal MRI at 3.0 T: relaxation times and image contrast.

Authors:  Garry E Gold; Eric Han; Jeff Stainsby; Graham Wright; Jean Brittain; Christopher Beaulieu
Journal:  AJR Am J Roentgenol       Date:  2004-08       Impact factor: 3.959

3.  Effect of testosterone replacement on trabecular architecture in hypogonadal men.

Authors:  Maria Benito; Branimir Vasilic; Felix W Wehrli; Benjamin Bunker; Michael Wald; Bryon Gomberg; Alexander C Wright; Babette Zemel; Andrew Cucchiara; Peter J Snyder
Journal:  J Bone Miner Res       Date:  2005-06-20       Impact factor: 6.741

4.  Phantom design method for high-field MRI human systems.

Authors:  Qing X Yang; Jinghua Wang; Christopher M Collins; Michael B Smith; Xiaoliang Zhang; Kamil Ugurbil; Wei Chen
Journal:  Magn Reson Med       Date:  2004-11       Impact factor: 4.668

5.  Improved data reconstruction method for GRAPPA.

Authors:  Ze Wang; Jiongjiong Wang; John A Detre
Journal:  Magn Reson Med       Date:  2005-09       Impact factor: 4.668

6.  Comparison of magnetic properties of MRI contrast media solutions at different magnetic field strengths.

Authors:  Martin Rohrer; Hans Bauer; Jan Mintorovitch; Martin Requardt; Hanns-Joachim Weinmann
Journal:  Invest Radiol       Date:  2005-11       Impact factor: 6.016

7.  Effects of salmon calcitonin on trabecular microarchitecture as determined by magnetic resonance imaging: results from the QUEST study.

Authors:  Charles H Chesnut; Sharmilla Majumdar; David C Newitt; Andrew Shields; Jan Van Pelt; Ellen Laschansky; Moise Azria; Audrey Kriegman; Melvin Olson; Erik F Eriksen; Linda Mindeholm
Journal:  J Bone Miner Res       Date:  2005-04-27       Impact factor: 6.741

8.  Accuracy of high-resolution in vivo micro magnetic resonance imaging for measurements of microstructural and mechanical properties of human distal tibial bone.

Authors:  X Sherry Liu; X Henry Zhang; Chamith S Rajapakse; Michael J Wald; Jeremy Magland; Kiranjit K Sekhon; Mark F Adam; Paul Sajda; Felix W Wehrli; X Edward Guo
Journal:  J Bone Miner Res       Date:  2010-09       Impact factor: 6.741

9.  Half-fourier acquisition single-shot turbo spin-echo (HASTE) MR: comparison with fast spin-echo MR in diseases of the brain.

Authors:  M R Patel; R A Klufas; R A Alberico; R R Edelman
Journal:  AJNR Am J Neuroradiol       Date:  1997-10       Impact factor: 3.825

10.  Quantitative high-resolution magnetic resonance imaging reveals structural implications of renal osteodystrophy on trabecular and cortical bone.

Authors:  Felix W Wehrli; Mary B Leonard; Punam K Saha; Bryon R Gomberg
Journal:  J Magn Reson Imaging       Date:  2004-07       Impact factor: 4.813

View more
  6 in total

1.  Evaluation of intervertebral disc cartilaginous endplate structure using magnetic resonance imaging.

Authors:  Sung M Moon; Jonathon H Yoder; Alexander C Wright; Lachlan J Smith; Edward J Vresilovic; Dawn M Elliott
Journal:  Eur Spine J       Date:  2013-05-15       Impact factor: 3.134

2.  Theory of MRI contrast in the annulus fibrosus of the intervertebral disc.

Authors:  Alexander C Wright; Jonathon H Yoder; Edward J Vresilovic; Dawn M Elliott
Journal:  MAGMA       Date:  2016-01-11       Impact factor: 2.310

3.  Internal three-dimensional strains in human intervertebral discs under axial compression quantified noninvasively by magnetic resonance imaging and image registration.

Authors:  Jonathon H Yoder; John M Peloquin; Gang Song; Nick J Tustison; Sung M Moon; Alexander C Wright; Edward J Vresilovic; James C Gee; Dawn M Elliott
Journal:  J Biomech Eng       Date:  2014-11       Impact factor: 2.097

4.  Influence of bone marrow composition on measurements of trabecular microstructure using decay due to diffusion in the internal field MRI: simulations and clinical studies.

Authors:  Sara M Sprinkhuizen; Jerome L Ackerman; Yi-Qiao Song
Journal:  Magn Reson Med       Date:  2013-12-31       Impact factor: 4.668

5.  A 16-Channel 13C Array Coil for Magnetic Resonance Spectroscopy of the Breast at 7T.

Authors:  Matthew Wilcox; Stephen Ogier; Sergey Cheshkov; Ivan Dimitrov; Craig Malloy; Steven Wright; Mary McDougall
Journal:  IEEE Trans Biomed Eng       Date:  2021-05-21       Impact factor: 4.756

Review 6.  RF coils: A practical guide for nonphysicists.

Authors:  Bernhard Gruber; Martijn Froeling; Tim Leiner; Dennis W J Klomp
Journal:  J Magn Reson Imaging       Date:  2018-06-13       Impact factor: 4.813

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.