Anterior capsulotomy with an ultrashort-pulse laser.

Ramon Naranjo Tackman1, Jorge Villar Kuri, Louis D Skip Nichamin, Keith Edwards.   

Abstract

PURPOSE: To assess the precision of laser anterior capsulotomy compared with that of manual continuous curvilinear capsulorhexis (CCC).
SETTING: Asociación Para Evitar La Ceguera en México IAP, Hospital Dr. Luis Sánchez Bulnes, Mexico City, Mexico.
DESIGN: Nonrandomized single-center clinical trial.
METHODS: In patients presenting for cataract surgery, the LensAR Laser System was used to create a laser anterior capsulotomy of the surgeon's desired size. Capsule buttons were retrieved and measured and then compared with buttons retrieved from eyes having a manually torn CCC. Deviation from the intended diameter and the regularity of shape were assessed. When removing the capsule buttons at the start of surgery, the surgeon rated the ease of removal on a scale of 1 to 10 (1 = required manual capsulorhexis around the whole diameter; 10 = button free floating or required no manual detachment from remaining capsule during removal).
RESULTS: The mean deviation from the intended diameter was 0.16 mm ± 0.17 (SD) for laser anterior capsulotomy and 0.42 ± 0.54 mm for CCC (P=.03). The mean absolute deviation from the intended diameter was 0.20 ± 0.12 mm and 0.49 ± 0.47 mm, respectively (P=.003). The mean of the average squared residuals was 0.01 ± 0.03 and 0.02 ± 0.04, respectively (P=.09). The median rating of the ease of removal was 9 (range 5 to 10).
CONCLUSIONS: Laser anterior capsulotomy created a more precise capsule opening than CCC, and the buttons created by the laser procedure were easy to remove at the beginning of cataract surgery.
Copyright © 2011 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21397456     DOI: 10.1016/j.jcrs.2010.11.030

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  8 in total

Review 1.  The journey to femtosecond laser-assisted cataract surgery: new beginnings or a false dawn?

Authors:  S Trikha; A M J Turnbull; R J Morris; D F Anderson; P Hossain
Journal:  Eye (Lond)       Date:  2013-02-01       Impact factor: 3.775

2.  Femtosecond laser capsulotomy versus manual capsulotomy: a Meta-analysis.

Authors:  Dao-Wei Qian; Hai-Ke Guo; Shang-Li Jin; Hong-Yang Zhang; Yuan-Cun Li
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

3.  Femtosecond laser-assisted cataract surgery in vitrectomized eye with posterior chamber phakic intraocular lens.

Authors:  Natalia Anisimova; Boris Malyugin; Lisa B Arbisser; Nikolay Sobolev
Journal:  Digit J Ophthalmol       Date:  2017-06-30

4.  Femtosecond cataract surgery: A review of current literature and the experience from an initial installation.

Authors:  Chris Hodge; Shveta Jindal Bali; Michael Lawless; Colin Chan; Timothy Roberts; David Ng; Simon Chen; Paul Hughes; Gerard Sutton
Journal:  Saudi J Ophthalmol       Date:  2012-01

5.  First experience with the new high-frequency femtosecond laser system (LDV Z8) for cataract surgery.

Authors:  Bojan Pajic; Iraklis Vastardis; Zisis Gatzioufas; Brigitte Pajic-Eggspuehler
Journal:  Clin Ophthalmol       Date:  2014-12-08

6.  Learning curve of femtosecond laser-assisted cataract surgery: Experience of surgeons new to femtosecond laser platform.

Authors:  Josephine Susai Christy; Manas Nath; Fredrick Mouttapa; Rengaraj Venkatesh
Journal:  Indian J Ophthalmol       Date:  2017-08       Impact factor: 1.848

Review 7.  Advances in femtosecond laser technology.

Authors:  Thais Pinheiro Callou; Renato Garcia; Adriana Mukai; Natalia T Giacomin; Rodrigo Guimarães de Souza; Samir J Bechara
Journal:  Clin Ophthalmol       Date:  2016-04-19

8.  Efficacy of 0.015% intracameral epinephrine for significant miosis induced by photodisruption during femtosecond laser-assisted cataract surgery.

Authors:  Jong Hwa Jun; Seung Pil Bang; Young-Sik Yoo; Choun-Ki Joo
Journal:  Medicine (Baltimore)       Date:  2018-08       Impact factor: 1.817

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.