| Literature DB >> 21388738 |
Shun-Xing Li1, Feng-Ying Zheng, Shu-Jie Cai, Tian-Shou Cai.
Abstract
The nanometer TiO(2) particle was coated onto the inner wall of a T-shaped quartz tube atomizer (QTA) and then was used as a new atomizer (NT-QTA) for the determination of Hg and Se by hydride generation atomic absorption spectrometry (HGAAS). After coating 67.4 mg TiO(2) on a quartz tube, the analytical performance of NT-QTA-HGAAS was compared to conventional QTA-HGAAS and it was improved as follows: (a) the linear range of the calibration curves was expanded from 10.0-80.0 ng mL(-1) to 5.0-150.0 ng mL(-1) for Hg, and from 10.0-70.0 ng mL(-1) to 5.0-100.0 ng mL(-1) for Se; (b) the characteristic concentration of was decreased from 2.8 ng mL(-1)/1% to 1.1 ng mL(-1)/1% for Hg and from 1.2 ng mL(-1)/1% to 0.8 ng mL(-1)/1% for Se; and (c) the interference from the coexistence of As on the determination of Hg and Se could be eliminated. The achieved technique was applied for the determination of Hg and Se in herbal medicines and hair.Entities:
Mesh:
Substances:
Year: 2011 PMID: 21388738 DOI: 10.1016/j.jhazmat.2011.02.019
Source DB: PubMed Journal: J Hazard Mater ISSN: 0304-3894 Impact factor: 10.588