Literature DB >> 21371913

Visual grading characteristics (VGC) analysis of diagnostic image quality for high resolution 3 Tesla MRI volumetry of the olfactory bulb.

Hartmut Peter Burmeister1, Pascal Andreas Thomas Baltzer, Constanze Möslein, Thomas Bitter, Hilmar Gudziol, Matthias Dietzel, Orlando Guntinas-Lichius, Werner Alois Kaiser.   

Abstract

RATIONALE AND
OBJECTIVES: The aim of this study was to evaluate the suitability of 3-T magnetic resonance imaging (MRI) for olfactory bulb volumetry, comparing image quality obtained using different sequences on the basis of physical characteristics in combination with observer performance.
MATERIALS AND METHODS: Twenty-two healthy volunteers (11 men, 11 women; mean age, 25 years) underwent 3-T MRI of the frontal skull base in this prospective study. Imaging was performed using a constructive interference in steady state (CISS) three-dimensional Fourier transformation sequence, a three-dimensional T2-weighted (3D-T2w) sequence, and a two-dimensional T2-weighted (2D-T2w) sequence. The relative performance of sequences was assessed using visual grading characteristic analysis. Interobserver agreement was assessed using κ statistics. For evaluation of physical image quality characteristics, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated and compared using Wilcoxon's test. SNR and CNR measurements were correlated with visual grading results.
RESULTS: Visual grading characteristic analysis showed significantly better image quality ratings for the CISS sequence compared to the 3D-T2w and 2D-T2w sequence, and the 2D-T2w sequence performed significantly better compared to the 3D-T2w sequence (P < .001). Interobserver agreement was substantial (κ = 0.66-0.73). Wilcoxon's test revealed significantly higher SNR and CNR values for the 2D-T2w sequence compared to the 3D-T2w and CISS sequences, and SNR and CNR values for the 3D-T2w sequence were significantly higher compared to those for the CISS sequence (P < .001 for each). Significant correlation between SNR and CNR and visual grading was found for the 3D-T2w sequence (SNR: ρ = 0.510, P = .015; CNR: ρ = 0.546, P = .009).
CONCLUSIONS: High-resolution 3-T MRI resulted in excellent values for SNR and CNR, suggesting the appropriateness of the sequences for olfactory bulb MRI volumetry. Visual grading characteristic analysis revealed the CISS sequence to be the most suitable for olfactory bulb volumetry.
Copyright © 2011 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2011        PMID: 21371913     DOI: 10.1016/j.acra.2010.12.011

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  Evaluation of a new system for chest tomosynthesis: aspects of image quality of different protocols determined using an anthropomorphic phantom.

Authors:  M Jadidi; A Sundin; P Aspelin; M Båth; S Nyrén
Journal:  Br J Radiol       Date:  2015-06-29       Impact factor: 3.039

2.  Dependency of image quality on acquisition protocol and image processing in chest tomosynthesis-a visual grading study based on clinical data.

Authors:  Masoud Jadidi; Magnus Båth; Sven Nyrén
Journal:  Br J Radiol       Date:  2018-04-09       Impact factor: 3.039

3.  A retrospective, semi-quantitative image quality analysis of cone beam computed tomography (CBCT) and MSCT in the diagnosis of distal radius fractures.

Authors:  H Lang; J Neubauer; B Fritz; E M Spira; J Strube; M Langer; E Kotter
Journal:  Eur Radiol       Date:  2016-03-22       Impact factor: 5.315

  3 in total