Sensitivity analysis of a geometric calibration method using projection matrices for digital tomosynthesis systems.

Xinhua Li1, Zhang Da, Bob Liu.   

Abstract

PURPOSE: To study the sensitivity of a geometric calibration method using projection matrices for digital tomosynthesis systems.
METHODS: A generic geometric calibration method for tomographic imaging systems has been presented in our previous work. The method involves a scan of a calibration phantom with multiple markers. Their locations in projection images are detected and are associated with their 3D coordinates to compute 3 x 4 projection matrices, which can be used in subsequent image reconstruction. The accuracy of geometric calibration may be affected by errors in the input data of marker positions. The effects of errors may depend on the number of markers and the volume surrounded by them in 3D space. This work analyzed the sensitivity of the calibration method to the above factors. A 6 cm CIRS breast research phantom and a prototype breast tomosynthesis system were used for our tests. A high contrast ring and two small speck groups were reconstructed in various testing cases for comparison. To achieve quantitative assessment, a 15 x 15 point detection mask was adopted for detecting signals and for computing changes between testing cases and the regular geometric calibration.
RESULTS: When 3D coordinates and 2D projections of markers were accurate, all tested numbers of markers, 6-44, provided similar high quality reconstructions of the ring and the two speck groups. Errors in marker positions resulted in image degradations and signal changes, which increased with fewer markers and smaller volume surrounded by markers in the 3D object space. Signal changes of small specks were more significant than those of the ring. Errors in marker projections produced drastic image degradations. Coplanar marker placement caused a failure in projection matrix computation.
CONCLUSIONS: For practical geometric calibration phantom design, ample markers are desired. They need to have a large volumetric coverage in the 3D space and be far from being coplanar. Precise determination of marker projections on detector planes is crucial for accurate geometric calibration and for small object detection using the reconstructed images.

Mesh:

Year:  2011        PMID: 21361188     DOI: 10.1118/1.3524221

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

Review 1.  A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications.

Authors:  Ioannis Sechopoulos
Journal:  Med Phys       Date:  2013-01       Impact factor: 4.071

2.  Digital breast tomosynthesis: observer performance of clustered microcalcification detection on breast phantom images acquired with an experimental system using variable scan angles, angular increments, and number of projection views.

Authors:  Heang-Ping Chan; Mitchell M Goodsitt; Mark A Helvie; Scott Zelakiewicz; Andrea Schmitz; Mitra Noroozian; Chintana Paramagul; Marilyn A Roubidoux; Alexis V Nees; Colleen H Neal; Paul Carson; Yao Lu; Lubomir Hadjiiski; Jun Wei
Journal:  Radiology       Date:  2014-07-07       Impact factor: 11.105

3.  Self-calibration of cone-beam CT geometry using 3D-2D image registration.

Authors:  S Ouadah; J W Stayman; G J Gang; T Ehtiati; J H Siewerdsen
Journal:  Phys Med Biol       Date:  2016-03-10       Impact factor: 3.609

4.  A geometric calibration method for the digital chest tomosynthesis with dual-axis scanning geometry.

Authors:  Chia-Hao Chang; Yu-Ching Ni; Syuan-Ya Huang; Ho-Hui Hsieh; Sheng-Pin Tseng; Fan-Pin Tseng
Journal:  PLoS One       Date:  2019-04-25       Impact factor: 3.240

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.