Evaluation of the interference due to haemoglobin, bilirubin and lipids on Immulite 2500 assays: a practical approach.

Gerard Steen1, Alex Klerk, Kees van der Laan, Elaine F Eppens.   

Abstract

BACKGROUND: Interfering substances such as haemoglobin, bilirubin and lipids in a sample may lead to wrong interpretation of immunoassay results by the clinician. In general, there has been minor attention to these interferences on immunoassays, whereas these effects on chemical assays are frequently described. Information about interferences by haemoglobin, bilirubin and lipids on the Siemens Immulite 2500 assays in the instructions for use is falling short.
METHODS: Interferents in patient samples can be measured reliably in a semi-quantitative way on most chemistry analysers and can be expressed in haemolysis-, icterus- and lipaemia-indices. As the Immulite 2500 cannot perform such measurements, samples are normally analysed without testing for the presence of interferents. Therefore, a study was carried out to examine these interferences on 24 Immulite 2500 assays. Samples were spiked with increasing concentrations of either haemoglobin, bilirubin or lipids. The haemolysis-, icterus- and lipaemia-indices were measured on a Synchron DxC 800 analyser.
RESULTS: Based on analytical imprecision and intraindividual biological variation of each analyte, cut-off indices above which clinically significant interference exists were determined. We found clinically significant interference due to haemoglobin on ferritin and folate, by bilirubin on oestradiol and testosterone and by lipids on testosterone.
CONCLUSIONS: Introducing cut-off indices prevents reporting of wrong Immulite 2500 results due to interference. Our results are applicable in laboratories using any chemistry analyser capable of reporting semi-quantitative concentrations of interferents.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 21355012     DOI: 10.1258/acb.2010.010187

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  6 in total

1.  Temporal kinetics of bovine mammary IgG secretion into colostrum and transition milk.

Authors:  Kasey M Schalich; Olivia M Reiff; Blake T Nguyen; Cassandra L Lamb; Cecilia R Mondoza; Vimal Selvaraj
Journal:  J Anim Sci       Date:  2021-05-01       Impact factor: 3.159

2.  Frequency of icteric interference in clinical chemistry laboratory tests and causes of severe icterus.

Authors:  Sandhya Mainali; Anna E Merrill; Matthew D Krasowski
Journal:  Pract Lab Med       Date:  2021-10-23

3.  Optimization of an ammonia assay based on transmembrane pH-gradient polymersomes.

Authors:  Anastasia Spyrogianni; Charlotte Gourmel; Leopold Hofmann; Jessica Marbach; Jean-Christophe Leroux
Journal:  Sci Rep       Date:  2021-11-11       Impact factor: 4.379

4.  Investigation of unusual high serum indices for lipemia in clear serum samples on siemens analysers dimension.

Authors:  Eva Fliser; Ksenija Jerkovic; Tanja Vidovic; Maksimiljan Gorenjak
Journal:  Biochem Med (Zagreb)       Date:  2012       Impact factor: 2.313

5.  Frequency and causes of lipemia interference of clinical chemistry laboratory tests.

Authors:  Sandhya Mainali; Scott R Davis; Matthew D Krasowski
Journal:  Pract Lab Med       Date:  2017-02-03

Review 6.  Performance and comparability of laboratory methods for measuring ferritin concentrations in human serum or plasma: A systematic review and meta-analysis.

Authors:  Maria N Garcia-Casal; Juan P Peña-Rosas; Eloisa Urrechaga; Jesus F Escanero; Junsheng Huo; Ricardo X Martinez; Lucero Lopez-Perez
Journal:  PLoS One       Date:  2018-05-03       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.