Abstract
INTRODUCTION: We present our work of monitoring 202 different patients with markedly elevated serum index for lipemia whereby serum samples were clear. We tried to clarify the cause of occurrence of these indices which were detected in the years 2006-2010 on Siemens Dimension analyzers.Entities:
Mesh:
Substances:
Year: 2012 PMID: 23092066 PMCID: PMC3900048 DOI: 10.11613/bm.2012.037
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Haemolysis, icterus and lipemia measurement wavelengths on Siemens Dimension, Roche Hitachi and former Olympus analyzers.
| Haemolysis | 405/700 | 570/600 + 660/700 | 410/480 + 600/800 |
| Icterus | 452/700 | 480/505 + 570/600 | 480/570 + 600/800 |
| Lipemia | 700 | 660/700 | 660/800 |
Siemens Dimension analysers – RXL, RXL Max, Xpand, Xpand Plus, Vista;
BC – Beckman Coulter with former Olympus analysers.
Concentrations of haemoglobin, bilirubin and triglycerides with appropriate serum indices for RXL Max and Vista analysers, provided by Siemens Healthcare Diagnostics.
| H≤0.25 | H≤0.10 | I≤34.2 | I≤34.2 | L≤0.28 | L≤0.56 | |
| 0.25<H≤0.50 | 0.10<H≤0.25 | 34.2<I≤85.5 | 34.2<I≤85.5 | 0.28<L≤0.57 | 0.56<L≤1.13 | |
| 0.50<H≤2.00 | 0.25<H 0.50 | 85.5<I≤342 | 85.5<I≤171 | 0.57<L≤2.26 | 1.13<L≤2.26 | |
| 2.00<H≤3.00 | 0.50<H≤2.00 | 342<I≤684 | 171<I≤256 | 2.26<L≤6.78 | 2.26<L≤4.52 | |
| 3.00<H≤5.00 | 2.00<H≤3.00 | 684<I≤1026 | 256<I≤342 | 6.78<L≤11.3 | 4.52<L≤6.78 | |
| 5.00<H≤10.0 | 3.00<H≤5.00 | 1026<I≤1368 | 342<I≤684 | 11.3<L≤33.9 | 6.78<L≤9.04 | |
| 5.00<H≤10.0 | 684<I≤1026 | 9.04<L≤11.3 | ||||
| H>10.0 | I>1026 | L>11.3 | ||||
Triglyceride concentration was simulated with Intralipid™ supplementation.
Number of patients with lipemia index from 4 to 6 (RXL Max) or from 4 to 8 (Vista) and a clear serum; sex and age of patients.
| Number of patients | 202 | 109 | 93 | |
| Women | 107 (53%) | 56 (51%) | 51 (55%) | |
| Age (years) | 70 (15–92) | 70 (19–91) | 70 (15–92) | |
| Mean serum index for lipemia | 4.3 ± 0.8 | 4.1 ± 0.4 | 4.6 ± 1.0 | < 0.001 |
Serum indices are presented as mean values to one decimal place for a better illustration of the differences in measurements.
P value for differences between mean values of serum indices on RXL Max and Vista analysers, Mann-Whitney Test.
Total number of samples with individual (monoclonal, biclonal or polyclonal) immunoglobulins and median concentration of relevant immunoglobulins (IgG, IgM, IgA); number of samples according to the lipemia index.
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| IgM + kappa | 67 | 14.80 (9.41–22.83) | 33 | 7 | - | 20 | 4 | 3 | - | - |
| IgG + kappa | 50 | 19.0 (13.7–32.5) | 24 | 4 | - | 12 | 3 | 3 | 2 | 2 |
| IgM + lambda | 29 | 25.00 (12.30–32.10) | 10 | 2 | - | 11 | 4 | - | 1 | 1 |
| IgG + lambda | 14 | 16.1 (11.5–23.8) | 6 | - | 1 | 7 | - | - | - | - |
| polyclonal Ig | 26 | 26.6 (18.5–30.5) | 13 | - | - | 12 | 1 | - | - | - |
| biclonal Ig | 15 | 20.6 (16.7–34.3) | 8 | - | - | 3 | 3 | - | - | 1 |
| IgA + lambda | 1 | 48.91 | 1 | - | - | - | - | - | - | |
Ig – immunoglobulins.
Median and interquartile range for IgM immunoglobulin of the monoclonal peak.
Median and interquartile range for IgG immunoglobulin of the monoclonal peak.
Median and interquartile range for IgG immunoglobulin of the polyclonal peak.
Median and interquartile range for IgG immunoglobulin of the biclonal peak.
Median and interquartile range for IgM immunoglobulin of the biclonal peak.
Concentration of the paraprotein.
Mean and median concentration values for individual tests that were performed to determine their impact on the phenomenon of high serum indices for lipemia in clear serum samples; values for 202 samples with unusual high lipemia index in a clear serum and values for the control group.
| TP (g/L) | 77 ± 14 | 76 ± 14 | 0.625 |
| CHOL (mmol/L) | 4.10 ± 1.25 | 4.00 ± 1.31 | 0.640 |
| TG (mmol/L) | 1.20 (0.88–1.62) | 1.43 (1.14–2.14) | 0.506 |
| HDL (mmol/L) | 1.09 ± 0.43 | 1.09 ± 0.37 | 0.990 |
| LDL (mmol/L) | 2.40 ± 0.90 | 2.33 ± 0.85 | 0.600 |
| ApoA1 (g/L) | 1.32 ± 0.53 | 1.25 ± 0.53 | 0.368 |
| ApoB (g/L) | 0.86 (0.58–1.18) | 0.74 (0.63–1.21) | 0.440 |
| Lp(a) (g/L) | <0.02–2.13 | <0.02–1.84 | 0.540 |
| RF (IU/mL) | <10–7990 | <10–903 | 0.688 |
TP - Total protein, CHOL – Total cholesterol, TG – Triglycerides, HDL – HDL cholesterol, LDL – LDL cholesterol, RF – Rheumatoid factor.
Mann-Whitney U-test,
Two-tailed T-test,
Two-tailed Pearson chi-square test (χ2).
Figure 1.Measurements of lipemia index in 42 samples of different patients with unusual high lipemia index on RXL Max, Vista, Hitachi 911 and former Olympus AU640 analyzers.
Parameters that were measured to evaluate their impact on the occurrence of high serum indices in clear serum samples; methods, measuring ranges, instruments and reagent manufacturers.
| IgG | Siemens BN II nephelometer | Nephelometric method | 1.4–64.0 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| IgA | Siemens BN II nephelometer | Siemens BN II nephelometer Nephelometric method | Siemens BN II nephelometer Nephelometric method 0.23–8.00 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| IgM | IgM Siemens BN II nephelometer | IgM Siemens BN II nephelometer Nephelometric method | 0.18–6.40 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| Kappa light chains | Kappa light chains Siemens BN II nephelometer | Nephelometric method | 0.28–9.10 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| Lambda light chains | Siemens BN II nephelometer | Nephelometric method | 0.16–5.00 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| Monoclonal immunoglobulin | Monoclonal immunoglobulin Interlab Microtech 648 PC | Monoclonal immunoglobulin Interlab Microtech 648 PC Serum protein electrophoresis | / | Interlab Scientific Instruments, Roma, Italy |
| Monoclonal immunoglobulin | Sebia Hydrasis | Immuno-electrophoresis | / | Sebia, Lisses, Cedex, France |
| Total cholesterol | Total cholesterol Siemens Dimension RXL Max, Vista | Cholesterol esterase | 1.30–15.50 mmol/L | Siemens HealthCare Diagnostics, Newark, DE, USA |
| Triglycerides | Siemens Dimension RXL Max, Vista | GPO-PAP Enzymatic | 0.02–11.30 mmol/L | Siemens HealthCare Diagnostics, Newark, DE, USA |
| HDL cholesterol | Siemens Dimension RXL Max, Vista | Homogenous direct (enzymatic colorimetric) method | 0.05–3.89 mmol/L | Siemens HealthCare Diagnostics, Newark, DE, USA |
| LDL cholesterol | Siemens Dimension RXL Max, Vista | Direct (enzymatic colorimetric) method | 0.00–7.80 mmol/L | Siemens HealthCare Diagnostics, Newark, DE, USA |
| Total protein (TP) | Siemens Dimension RXL Max, Vista | Photometric colour test (biuret) | 0–120 g/L | Siemens HealthCare Diagnostics, Newark, DE, USA |
| Lp(a) | Siemens BN ProSpec nephelometer | Nephelometric method | 0.10–1.60 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| ApoA1 | Siemens BN II nephelometer | Nephelometric method | 0.19–6.00 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| ApoB | Siemens BN II nephelometer | Nephelometric method | 0.25–4.00 g/L | Siemens HealthCare Diagnostics Products GmbH, Marburg, Germany |
| Rheumatoid factor (RF) | Siemens BN ProSpec nephelometer | Nephelometric method | 10–600 IU/mL | Siemens HealthCare Diagnostics Products GmbH, Marburg, |
| Cryoglobulins | PU 8675 VIS Spectrophotometer | Lowry method | 0–400 mg/L | Germany Philips, Cambridge, United Kingdom |
| Serum index | Olympus AU640 | LIH | / | Olympus Diagnostica GmbH, Lismeehan, Ireland |
| Serum index | Hitachi 911 | LIH | / | Roche Diagnostics GmbH, Mannheim, Germany |
| Serum index | Siemens Dimension RXL Max, Vista | HIL | / | Siemens HealthCare Diagnostics, Newark, DE, USA |
A change of lower and upper range value is possible with dilution of the sample.
A change of upper range value is possible with dilution of the sample.
Detection limit for monoclonal immunoglobulin is 1.0 g/L.
Detection limit for monoclonal immunoglobulin is between 0.12–0.25 g/L.
Name regarding the time of measurements; otherwise Beckman Coulter.