The effect of catheter to vein ratio on blood flow rates in a simulated model of peripherally inserted central venous catheters.

Thomas P Nifong1, Timothy J McDevitt2.   

Abstract

BACKGROUND: Catheter-related thrombosis is a common complication in all anatomic sites, especially when smaller veins of the upper extremity are considered. It is presumed that the presence of a catheter within the lumen of a vein will decrease flow and potentially create stasis, and clinical data suggest that the size of the catheter impacts thrombosis rates. We sought to determine, both mathematically and experimentally, the impact of catheters on fluid flow rates.
METHODS: We used fluid mechanics to calculate relative flow rates as a function of the ratio of the catheter to vein diameters. We also measured the flow rate of a blood analyte solution in an annular flow model using diameters that simulate the size of upper extremity veins and commonly used peripherally inserted central catheters (PICCs).
RESULTS: We compared each of the derived relative flow rates to the experimentally determined ones for three cylinder sizes and found a correlation of r(2) = 0.90. We also confirmed that the decrease in fluid flow rate with each successive catheter size is statistically significant (P < .0001).
CONCLUSIONS: Our results demonstrate that fluid flow is dramatically decreased by the insertion of a centrally located obstruction. Assuming that blood flow in veins behaves in a similar manner to our models, PICCs, in particular, may substantially decrease venous flow rates by as much as 93%.

Entities:  

Mesh:

Year:  2011        PMID: 21349931     DOI: 10.1378/chest.10-2637

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  33 in total

Review 1.  International evidence-based recommendations on ultrasound-guided vascular access.

Authors:  Massimo Lamperti; Andrew R Bodenham; Mauro Pittiruti; Michael Blaivas; John G Augoustides; Mahmoud Elbarbary; Thierry Pirotte; Dimitrios Karakitsos; Jack Ledonne; Stephanie Doniger; Giancarlo Scoppettuolo; David Feller-Kopman; Wolfram Schummer; Roberto Biffi; Eric Desruennes; Lawrence A Melniker; Susan T Verghese
Journal:  Intensive Care Med       Date:  2012-05-22       Impact factor: 17.440

2.  Totally Implantable Venous Access Devices: Efforts Are Needed to Standardize Procedures to Avoid Complication: Reply.

Authors:  Daniele Biacchi; Paolo Sammartino; Angelo Di Giorgio
Journal:  World J Surg       Date:  2016-07       Impact factor: 3.352

Review 3.  Reducing catheter-related thrombosis using a risk reduction tool centered on catheter to vessel ratio.

Authors:  Timothy R Spencer; Keegan J Mahoney
Journal:  J Thromb Thrombolysis       Date:  2017-11       Impact factor: 2.300

4.  Association between prior peripherally inserted central catheters and lack of functioning arteriovenous fistulas: a case-control study in hemodialysis patients.

Authors:  Mireille El Ters; Gregory J Schears; Sandra J Taler; Amy W Williams; Robert C Albright; Bernice M Jenson; Amy L Mahon; Andrew H Stockland; Sanjay Misra; Scott L Nyberg; Andrew D Rule; Marie C Hogan
Journal:  Am J Kidney Dis       Date:  2012-06-15       Impact factor: 8.860

5.  Ultrasound-guided radial arterial cannulation: long axis/in-plane versus short axis/out-of-plane approaches?

Authors:  Derya Berk; Yavuz Gurkan; Alparslan Kus; Halim Ulugol; Mine Solak; Kamil Toker
Journal:  J Clin Monit Comput       Date:  2013-02-16       Impact factor: 2.502

Review 6.  Central venous catheter-related thrombosis and thromboprophylaxis in children: a systematic review and meta-analysis.

Authors:  E Vidal; A Sharathkumar; J Glover; E V S Faustino
Journal:  J Thromb Haemost       Date:  2014-06-19       Impact factor: 5.824

7.  Distance of the internal central venous catheter tip from the right atrium is positively correlated with central venous thrombosis.

Authors:  David H Ballard; Navdeep S Samra; Karen Mathiesen Gifford; Robert Roller; Bruce M Wolfe; John T Owings
Journal:  Emerg Radiol       Date:  2016-04-25

8.  The impact of the catheter to vein ratio on peripheral intravenous cannulation success, a post-hoc analyses.

Authors:  Fredericus H J van Loon; Hendrikus H M Korsten; Angelique T M Dierick-van Daele; Arthur R A Bouwman
Journal:  PLoS One       Date:  2021-05-24       Impact factor: 3.240

9.  Realistic boundary conditions in SimVascular through inlet catheter modeling.

Authors:  Amirtahà Taebi; Selin Berk; Emilie Roncali
Journal:  BMC Res Notes       Date:  2021-05-31

10.  Power injectable peripherally inserted central catheters in the ICU: not only a question of flow.

Authors:  Manuela Bonizzoli; Stefano Batacchi; Giovanni Zagli; Adriano Peris
Journal:  Crit Care       Date:  2012-12-12       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.