| Literature DB >> 34029356 |
Fredericus H J van Loon1,2, Hendrikus H M Korsten3, Angelique T M Dierick-van Daele4,5, Arthur R A Bouwman3,1.
Abstract
BACKGROUND: Intravenous cannulation is usually the first procedure performed in modern healthcare, although establishing peripheral intravenous access is challenging in some patients. The impact of the ratio between venous diameter and the size of the inserted catheter (catheter to vein ratio, CVR) on the first attempt success rate can be of added value in clinical. This study tries to give insight into the consideration that must be made when selecting the target vein and the type of catheter, and proved the null hypothesis that an optimal CVR would not be associated with increased first attempt cannulation success.Entities:
Year: 2021 PMID: 34029356 PMCID: PMC8143382 DOI: 10.1371/journal.pone.0252166
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Demographics of the included patients.
| Variable | Value | |
|---|---|---|
| Sex | 342 (56%) | |
| 268 (44%) | ||
| Age | 55 ±14 | |
| BMI | 28 ±5 | |
| Number of attempts to successful cannulation | 506 (83%) | |
| 67 (11%) | ||
| 19 (3%) | ||
| 18 (3%) |
Relevant outcomes regarding the different study groups, based on the size of the inserted catheter.
| Size of the inserted catheter | Success rate on the first attempt | Catheter to vein ratio | Venous diameter |
|---|---|---|---|
| 22 gauge | 50 (66%) | 0.50 (0.33) | 2.0 (0.9) |
| 20 gauge | 113 (77%) | 0.46 (0.27) | 2.6 (1.0) |
| 18 gauge | 208 (90%) | 0.38 (0.18) | 3.4 (1.0) |
| 16 gauge | 76 (89%) | 0.40 (0.23) | 4.3 (1.2) |
| 14 gauge | 61 (86%) | 0.43 (0.14) | 4.8 (0.7) |
Success rate on the first attempt is represented as proportions (percentages). Catheter to vein ratio and venous diameter in millimeters are represented as median (interquartile range).
Fig 1ROC curve to determine the optimal cut-off point of the CVR.
ROC curve regarding the CVR, based on the outcome of interest, with an AUC of 73% (68% to 77%) and optimal cut-off point of 0.41 (100-specificity 0.26 and sensitivity 0.67).
Fig 2The relation between venous diameter and the size of the inserted catheter to guarantee a CVR<41%.
Venous diameter in millimeters (vertical axes) and the size of the inserted catheter (horizontal axes).