Hand-held dynamic contour tonometry.

Pascal B Knecht1, Ursina Schmid, Timo Romppainen, Annette Hediger, Jens Funk, Hartmut Kanngiesser, Christoph Kniestedt.   

Abstract

PURPOSE: We present a prototype of the hand-held dynamic contour tonometer (HH-DCT) and prospectively compare this HH-DCT with the well-established Perkins applanation tonometer (PAT) and the TonoPenXL (TPXL).
METHODS: In a prospective, single-centre, randomized study, intraocular pressure (IOP) readings were taken in random order using HH-DCT, PAT and TPXL tonometers. Intra-observer variability was calculated for each observer and compared between three experienced ophthalmologists and an inexperienced medical student.
RESULTS: Ninety-two corneas of 92 healthy participants were enrolled. IOP [mean mmHg ± standard deviation (SD)] as measured by HH-DCT was 16.97 ± 2.71, by PAT 13.98 ± 2.52 and by TPXL 13.34 ± 2.68. The range of three consecutive IOP readings differed significantly between the devices [p < 0.001; mean range: 1.45 ± 1.07 (HH-DCT), 1.87 ± 0.97 (PAT) and 2.08 ± 1.77 (TPXL)]. There was no difference of the range in all devices between the ophthalmologists and the medical student (HH-DCT p = 0.68, PAT p = 0.54, TPXL p = 0.48).
CONCLUSION: IOP readings measured by HH-DCT are significantly higher than by PAT and TPXL. The differences of IOP measurements are in good accordance with previous studies using the slit-lamp-mounted DCT (SL-DCT) and Goldmann Applanation Tonometry, where SL-DCT readings were 1-3.2 mmHg higher. HH-DCT seems to give more constant results, which can be seen in the lower intra-observer variability compared to PAT and TPXL.
© 2009 The Authors. Journal compilation © 2009 Acta Ophthalmol.

Entities:  

Mesh:

Year:  2011        PMID: 21348963     DOI: 10.1111/j.1755-3768.2009.01625.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  1 in total

1.  [Reduction of intraocular pressure and ocular pulse amplitude during general anesthesia].

Authors:  H Zuche; E Morinello; A Viestenz; M Fiorentzis; B Seitz; T Volk; A Viestenz
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.